What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DC use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.
- Lot or code
- All Serial Numbers distributed between the dates of April 25, 2022 - July 8, 2025
- FDA recall number
- Z-2538-2025
- Quantity
- 25,389 units
Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCK use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.
- Lot or code
- All Serial Numbers distributed between the dates of April 25, 2022 - July 8, 2025
- FDA recall number
- Z-2539-2025
- Quantity
- 15,834 units
Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCMC use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.
- Lot or code
- All Serial Numbers distributed between the dates of April 25, 2022 - July 8, 2025
- FDA recall number
- Z-2540-2025
- Quantity
- 8,413 units
Distribution (FDA): US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: CAN, NZL, KOR, BMU, GBR, JPN, TWN, AUS, BRA, COL, MEX, ABW, SWE, CHL, ARG, HKG
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Max Mobility LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.