Skip to content
FDAMedical devicesClass IOngoing

FDA recall 97354: mo-Vis BVBA: Micro Joystick R-net. Electrical wheelchair component. and 5 more

FDA recall 97354 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

5 more codes are in the record below.

  • Micro Joystick R-net. Electrical wheelchair component.

    Lot or code
    Model No. P002-52; UDI-DI: 05407008320027; Serial No. 1000 to 1690.
    FDA recall number
    Z-2454-2025
    Quantity
    3 units
  • Multi Joystick R-net. Electrical wheelchair component.

    Lot or code
    Model No. P002-62; UDI-DI: 05407008320041; Serial No. 1000 to 1690.
    FDA recall number
    Z-2455-2025
    Quantity
    3 units
  • IDM-MICRO-R. Electrical wheelchair component.

    Lot or code
    Model No. PRSPS0014; UDI-DI: 05407008320836; Serial No. 1000 to 1690.
    FDA recall number
    Z-2456-2025
    Quantity
    15 units
  • IDM-MULTI-R. Electrical wheelchair component.

    Lot or code
    Model No. PRSPS0016; UDI-DI: 05407008320850; Serial No. 1000 to 1690.
    FDA recall number
    Z-2457-2025
    Quantity
    15 units
  • All-round Joystick R-net Light. Electrical wheelchair component.

    Lot or code
    Model No. P002-76; UDI-DI: 05407008320096; Serial No. 1000 to 2647.
    FDA recall number
    Z-2458-2025
    Quantity
    3 units
  • IDM-ARLITE-R. Electrical wheelchair component.

    Lot or code
    Model No. PRSPS00018. UDI-DI: 05407008320102; Serial No. 1000 to 2647.
    FDA recall number
    Z-2459-2025
    Quantity
    25 units

Distribution (FDA): US Nationwide distribution in the state of TX.

Source

Issued by
U.S. Food and Drug Administration
Firm
mo-Vis BVBA
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: September 2025, Week 2025-W37. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97354 on September 10, 2025.