Why it was recalled
The internal label affixed to the sample tip rack is incorrectly labeled as reagent tip racks. When scanned by the Agility instrument, the tip rack is misidentified as a different tip type, rendering it incompatible with the system and preventing use which causes a delay in obtaining patient results.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Dynex Agility Sample Tips; Model Number: 67910; The 67910 Agility Sample Tips are an accessory/consumable used with the Agility Automated ELISA system.
- Lot or code
- Model Number: 67910; UDI-DI: 5060456180072; Lot Number: 863366;
- FDA recall number
- Z-2345-2025
- Quantity
- 237 units
Distribution (FDA): Domestic: CA, FL, NC, NJ, NY;