What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
STA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop analyzer capable of simultaneously performing clotting, chromogenic, and immunologic assays.
- Lot or code
- Product Code: All references. UDI-DI: 03607450581041, 03607450581027. All software versions.
- FDA recall number
- Z-2473-2025
- Quantity
- 3,650 units (538 US)
STA-Liquid Anti-Xa 4. Model/Catalog Number: 00311US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins by measuring their anti-Xa activity on antithrombin in a competitive assay using a synthetic chromogenic substrate.
- Lot or code
- Model/Catalog Number: 00311US. UDI-DI: 03607450011784.
- FDA recall number
- Z-2474-2025
- Quantity
- 116,600 units (114,600 US)
STA-Liquid Anti-Xa 8. Model/Catalog Number: 00322US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins by measuring their anti-Xa activity on antithrombin in a competitive assay using a synthetic chromogenic substrate.
- Lot or code
- Model/Catalog Number: 00322US. UDI-DI: 03607450011791
- FDA recall number
- Z-2475-2025
- Quantity
- 126,200 units (125,900 US)
STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.
- Lot or code
- Model/Catalog Number: 00674. UDI-DI: 13607450006749
- FDA recall number
- Z-2476-2025
- Quantity
- 829,800 units (245,200 US)
Distribution (FDA): Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran, Israel, Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Oman, Paraguay, Peru, Republique Tcheque, Romania, Saudi Arabia, Serbia, South Africa, Switzerland, Taiwan, Thailand, Tunisia, Turquie, United Arab Emirates, Uruguay, Vietnam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Diagnostica Stago, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.