What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
391 more codes are in the record below.
Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion 3 M15, Model Number 722064. 3. Azurion 7 B12, Model Number 722067. 4. Azurion 7 B20, Model Number 722068. 5. Azurion 7 M12, Model Number 722078. 6. Azurion 7 M20, Model Number 722079. 7. Azurion 3 M12, Model Number 722221. 8. Azurion 3 M15, Model Number 722222. 9. Azurion 7 M12, Model Number 722223. 10. Azurion 7 M20, Model Number 722224. 11. Azurion 7 B12, Model Number 722225. 12. Azurion 7 B20, Model Number 722226. 13. Azurion 5 M12, Model Number 722227. 14. Azurion 5 M20, Model Number 722228. 15. Azurion 7 M20, Model Number 722282.
- Lot or code
- 1. Model Number/System Code (UDI); Serial Numbers: 722063 (00884838085275); 25, 151, 91, 85, 90, 86, 107, 89, 88, 87, 120, 153, 96, 108. 2. Model Number/System Code (UDI); Serial Numbers: 722064 (00884838085282); 64, 260, 261, 335, 218, 267, 270, 258, 35, 10, 111, 82, 125, 94, 278, 36, 55, 12, 57, 69, 484, 451, 497, 155, 272, 490, 132, 70, 251, 226, 290, 186, 48, 373. 3. Model Number/System Code (UDI); Serial Numbers: 722067 (00884838085350); 150, 69, 221, 148, 42, 98, 123, 277, 92, 103, 41, 214, 173, 172, 163, 128, 151, 22, 23, 268, 112, 97, 243, 45, 108, 113, 127, 176, 165, 33, 30, 206, 203, 48, 149, 251, 262. 4. Model Number/System Code (UDI); Serial Numbers: 722068 (00884838085367); 213, 111, 85, 89, 320, 73, 308, 172, 295, 406, 285, 413, 322, 324, 184, 442, 279, 361, 490, 42, 103, 520, 470, 472, 457, 400, 547, 510, 239, 275, 265, 365, 481, 62, 552, 44, 158, 36, 276, 68, 551, 516, 372, 46, 185, 515, 347, 416, 242, 219, 299, 126, 211, 468, 238, 568, 583, 69, 60, 426, 427, 93, 303, 302, 141. 5. Model Number/System Code (UDI); Serial Numbers: 722078 (00884838085251); 465, 143, 278, 402, 290, 524, 412, 166, 539, 532, 291, 415, 753, 441, 305, 289, 803, 399, 536, 537, 478, 479, 224, 163, 719, 364, 239, 383, 829, 513, 661, 215, 677, 709, 410, 269, 185, 902, 360, 546, 294, 906, 603, 870, 523, 598, 317, 535, 359, 436, 165, 733, 218, 204, 205, 206, 888, 923, 197, 195, 797, 444, 222, 489, 361, 362, 30, 223, 584, 962, 893, 27, 977, 156, 909, 953, 182, 958, 847, 959, 610, 490, 103,…
- FDA recall number
- Z-2502-2025
- Quantity
- 1,861 units
The text above is the beginning of the record; 9,373 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allura Xper FD10/10, Model Number 722005 3. Allura Xper FD20, Model Number 722006 4. Allura Xper FD20 Biplane, Model Number 722008 5. Allura Xper FD10, Model Number 722010 6. Allura Xper FD10/10, Model Number 722011 7. Allura Xper FD20, Model Number 722012 8. Allura Xper FD20 Biplane, Model Number 722013 9. Allura Xper FD20 OR Table, Model Number 722023 10. Allura Xper FD10, Model Number 722026 11. Allura Xper FD10/10, Model Number 722027 12. Allura Xper FD20, Model Number 722028 13. Allura Xper FD20/10, Model Number 722029 14. Allura Xper FD20 OR Table, Model Number 722035 15. Allura Xper FD20/20, Model Number 722038 16. Allura Xper FD20/15, Model Number 722058 17. Allura Xper FD20/15 OR Table, Model Number 722059.
- Lot or code
- 1. Model Number/System Code (UDI); Serial Numbers: 722003 (No UDI); 1225, 789, 1138, 445, 740, 152, 1411. 2. Model Number/System Code (UDI); Serial Numbers: 722005 (No UDI); 35. 3. Model Number/System Code (UDI); Serial Numbers: 722006 (No UDI); 168, 2198, 2184, 2606, 195, 161, 1956, 2256, 2070, 1502, 376, 1096, 1731. 4. Model Number/System Code (UDI); Serial Numbers: 722008 (No UDI); 246, 2470. 5. Model Number/System Code (UDI); Serial Numbers: 722010 (00884838059030); 911, 1180, 936, 338, 660, 411, 992, 1210, 873, 1028, 921. 6. Model Number/System Code (UDI); Serial Numbers: 722011 (00884838059047); 308. 7. Model Number/System Code (UDI); Serial Numbers: 722012 (00884838059054); 525, 823, 1141, 1078, 1811, 2092, 1641, 114, 2134, 1589, 1446, 1675, 1625, 1989, 2156, 455, 233, 2929, 2930, 749, 1080, 149, 854. 8. Model Number/System Code (UDI); Serial Numbers: 722013 (00884838059061); 580, 254, 78, 357, 428. 9. Model Number/System Code (UDI); Serial Numbers: 722023 (00884838059085); 164, 66. 10. Model Number/System Code (UDI); Serial Numbers: 722026 (00884838054189); 287, 825, 462, 192, 74, 191, 204, 155, 967, 1132, 205, 1246, 253, 156, 234, 233, 845, 122, 259, 185, 618, 531, 855, 235, 335, 1068, 843, 805, 646, 965, 258, 220, 602, 908, 221, 664, 109, 110, 116, 117, 115, 284, 389, 223, 694, 222, 436, 1192, 102, 914, 396, 704, 370, 201, 774, 696, 838, 344, 93, 1072, 601, 245, 1340, 604, 627, 190, 605, 935, 297, 98, 99, 100, 101, 544, 1129, 1268, 841, 1239, 1293, 648, 1038, 1231,…
- FDA recall number
- Z-2503-2025
- Quantity
- 1,006 units
The text above is the beginning of the record; 4,952 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Latvia, Lithuania, Macedonia, Maldives, Malta, Mexico, Moldova, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Panama, Paraguay, Peru, Philippines, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.