Skip to content
FDAMedical devicesClass IOngoing

FDA recall 97243: Boston Scientific Corporation: Carotid WALLSTENT Monorail Endoprosthesis Closed Cell Self-Expanding Stent

FDA recall 97243 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Certain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while attempting to withdraw the stent delivery system (SDS) from the guidewire or embolic protection device (EPD) after successful stent deployment. The most serious potential adverse health consequence is stroke. Other risks include delay to procedure, vessel injury, vessel spasm, or stent disruption/damage necessitating additional intervention.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • Carotid WALLSTENT Monorail Endoprosthesis Closed Cell Self-Expanding Stent, stent and stent delivery system. Material numbers: H7493915010240 CAROTID WALLSTENT MONORAIL 10.0-24; H7493915010310 CAROTID WALLSTENT MONORAIL 10.0-31; H749391506220 CAROTID WALLSTENT MONORAIL 6.0-22; H749391508210 CAROTID WALLSTENT MONORAIL 8.0-21; H749391508290 CAROTID WALLSTENT MONORAIL 8.0-29; H965SCH647010 CAROTID WALLSTENT MONORAIL 6.0-22; H965SCH647070 CAROTID WALLSTENT MONORAIL 8.0-21; H965SCH647080 CAROTID WALLSTENT MONORAIL 8.0-29; H965SCH647090 CAROTID WALLSTENT MONORAIL 8.0-36; H965SCH647120 CAROTID WALLSTENT MONORAIL 10.0-24; H965SCH647130 CAROTID WALLSTENT MONORAIL 10.0-31; H965SCH647140 CAROTID WALLSTENT MONORAIL 10.0-37; M001719000 CAROTID WALLSTENT MR 6X22 5F 135CM; M001719010 CAROTID WALLSTENT MR 8X21 5F 135CM; M001719020 CAROTID WALLSTENT MR 8X29 5F 135CM; M001719030 CAROTID WALLSTENT MR 8X36 5F 135CM; M001719040 CAROTID WALLSTENT MR 10X24 5.9F 135CM; M001719050 CAROTID WALLSTENT MR 10X31 5.9F 135CM; M001719060 CAROTID WALLSTENT MR 10X37 5.9F 135CM;

    Lot or code
    1. H7493915010240 UDI-DI 08714729781097 Batches 35277608 35315983 35315984 35315985 35380314 35452835 35452836 35514614 35551688 35551689 35664523 35664524 35750157 35750158 35750159 35750400 35879409 35879410 35921915 35922951 35926817 35926818 35942366 36056965 36058841 36058842 36058843 36096638 36121883 36138618 36138619 36138801 36138802 36138803 36159671 36195057 36195058 36195059 36195360 36205167 36209423 36214805 36214806 36256192 36322725 36348207 36348208 36348209 36498256 36588496 36607566 36711321; 2. H7493915010310 UDI-DI 08714729781110 Batches 35294379 35294540 35295922 35330503 35330504 35336132 35347263 35382292 35383351 35383352 35453002 35461487 35461488 35528579 35529200 35628964 35628965 35628966 35641995 35662047 35664484 35664485 35750401 35750402 35750403 35750404 35750405 35750406 35750407 35890012 35893584 35893585 35894358 35894359 35938922 35938923 35938924 35938925 35938926 36002641 36002642 36063291 36063292 36105446 36133161 36134808 36144148 36148535 36154091 36154092 36154093 36154094 36154095 36169871 36169872 36197809 36197810 36197811 36197812 36197813 36200118 36202762 36204350 36212466 36212467 36212468 36256193 36256194 36256195 36357711 36357712 36368143 36368144 36368145 36516036 36603362 36603363 36642784 36683543 36688775 36811476 36811477 36818953; 3. H749391506220 UDI-DI 08714729781158 Batches 35582211 35750408 35947814 35947815 35950476 36009906 36292862 36397124 36652844; 4. H749391508210 UDI-DI 08714729781172 Batches 35343574…
    FDA recall number
    Z-2273-2025
    Quantity
    26,570 (1,333 US; 25,174 OUS)

    The text above is the beginning of the record; 24,547 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Domestic US Nationwide distribution including Puerto Rico. International distribution worldwide.

Source

Issued by
U.S. Food and Drug Administration
Firm
Boston Scientific Corporation (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Boston Scientific Corporation, Month: August 2025, Week 2025-W34. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Certain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while attempting to withdraw the stent delivery system (SDS) from the guidewire or embolic protection device (EPD) after successful stent deployment. The most serious potential adverse health consequence is stroke. Other risks include delay to procedure, vessel injury, vessel spasm, or stent disruption/damage necessitating additional intervention.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97243 on August 20, 2025.

Other notices of Boston Scientific Corporation