What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
391 more codes are in the record below.
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410 Surgical Energy Platform (SEP). Model Number: WA91327U. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
- Lot or code
- Model Number: WA91327U. Lot Code: ESG-410 (SEP) WA91327U. UDI-DI: 04042761087698. Serial Numbers: 101696, 101697, 101698, 101699, 101700, 101701, 101702, 101703, 101704, 101705, 101706, 101707, 101708, 101709, 101710, 101919, 101920, 101921, 101922, 101923, 101924, 101925, 101926, 101927, 101928, 101929, 101933, 101934, 101935, 101936, 101937, 101940, 101941, 101942, 101945, 101946, 101947, 102148, 102149, 102150, 102151, 102152, 102153, 102154, 102156, 102157, 102158, 102159, 102160, 102161, 102162, 102164, 102165, 102167, 102168, 102170, 102172, 102173, 102174, 102175, 102176, 102177, 102178, 102179, 102180, 102181, 102182, 102183, 102184, 102185, 102186, 102187, 102188, 102189, 102190, 102191, 102193, 102194, 102195, 102198, 102199, 102200, 102266, 102427, 101932, 102155, 102163, 102197, 102660, 102661, 102662, 102663, 102664, 102665, 102666, 102667, 102668, 102669, 102670, 102671, 102672, 102673, 102674, 102675, 102676, 102677, 102678, 102679, 102680, 102681, 102682, 102683, 102684, 102685, 102686, 102687, 102688, 102689, 102690, 102691, 102692, 102693, 102694, 102695, 102696, 102697, 102698, 102699, 102700, 102701, 102702, 102703, 102704, 102705, 102706, 102707, 102708, 102709, 102710, 102711, 102712, 102713, 102714, 102715, 102716, 102717, 102718, 102719, 102721, 102722, 102723, 102724, 102725, 103922, 103930.
- FDA recall number
- Z-2338-2025
- Quantity
- 155 units
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open surgery, laparoscopic surgery, including single-site surgery, and endoscopic surgery.
- Lot or code
- Model Number: WA91307C. Lot Code: ESG-410 Legacy WA91307C. UDI-DI: 04042761086349. Serial Numbers: 101004, 101005, 101006, 101007, 101008, 101009, 101010, 101011, 101012, 101013, 101014, 101015, 101016, 101017, 101018, 101019, 101020, 101021, 101022, 101023, 101024, 101025, 101029, 101030, 101031, 101032, 101033, 101034, 101035, 101036, 101037, 101038, 101039, 101040, 101041, 101042, 101043, 101044, 101045, 101046, 101047, 101048, 101049, 101050, 101051, 101052, 101053, 101054, 101055, 101056, 101057, 101058, 101059, 101060, 101061, 101062, 101063, 101064, 101066, 101067, 101068, 101069, 101070, 101071, 101072, 101073, 101074, 101238, 101239, 101240, 101241, 101242, 101243, 101244, 101245, 101246, 101247, 101250, 101251, 101252, 101253, 101254, 101255, 101256, 101257, 101258, 101259, 101260, 101261, 101262, 101263, 101264, 101265, 101266, 101267, 101268, 101269, 101270, 101271, 101272, 101273, 101274, 101275, 101276, 101277, 101278, 101279, 101280, 101281, 101282, 101283, 101284, 101285, 101286, 101287, 101288, 101289, 101290, 101291, 101292, 101293, 101294, 101295, 101296, 101297, 101298, 101299, 101300, 101301, 101302, 101333, 101334, 101335, 101336, 101337, 101338, 101339, 101340, 101341, 101342, 101370, 101371, 101372, 101373, 101374, 101375, 101376, 101377, 101378, 101379, 101395, 101396, 101397, 101398, 101399, 101400, 101401, 101402, 101403, 101432, 101433, 101434, 101435, 101436, 101437, 101438, 101439, 101440, 101441, 101613, 101633, 101634, 101635, 101636, 101637,…
- FDA recall number
- Z-2339-2025
- Quantity
- 434 units
The text above is the beginning of the record; 2,072 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): US Nationwide distribution.