What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Low deaeration chamber level at post priming tests may lead to PrisMax System alarm T2309: Air Detected in Prime resulting in multiple additional priming cycles Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD Self-Test Failure Thermax System may be unable to detect the presence of a bag on the Thermax Blood Warmer Unit leading to difficulty setting up therapy, or PrisMax System alarm T2284: Thermax Disposable Not Inserted
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
PrisMax V2-US, Hemodialysis Delivery System, Product Codes: 955626, 955558, 955627, 955701
- Lot or code
- UDI 00085412639499, All Serial Numbers
- FDA recall number
- Z-2437-2025
- Quantity
- 4772 units
PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724
- Lot or code
- UDI 07332414126018, All serial numbers
- FDA recall number
- Z-2438-2025
- Quantity
- 2449 units
TherMax Blood Warmer Unit-US, PrisMax Accessory, Product Codes: 955630, 955515, 955702, 955631
- Lot or code
- UDI 07332414124731, All serial numbers
- FDA recall number
- Z-2439-2025
- Quantity
- 7432 units
Distribution (FDA): Worldwide
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- VANTIVE US HEALTHCARE LLC
- Title