Why it was recalled
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
- Lot or code
- Monitor UDI-DI: 00858481006017. Lot (C20206): 231220
- FDA recall number
- Z-2198-2025
- Quantity
- 2160
Distribution (FDA): US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.