What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR 1010.4 (Variances), 21 CFR 1010.2 (Certification), 21 CFR 1010.3 (Identification) and 21 CFR 1040.10 (g)(Labelling requirements). The identification label does not include the date of manufacture as required by 21 CFR 1010.3. The corrective action plan covers three families of laser light show of projectors: DIODE. CLUB and PURE with three different hardware configurations such as the number of the number of RJ45 ports, radiant power of emissions and network configurations. No specific user manuals were provided to each model family as required by 21 CFR 1040.10(h)(1) (Informational requirements (1) User information). The Quality Control test did not validate the manual reset function to enable resumption of laser radiation emission from Class IV laser projectors after an interruption caused using a remote interlock. No supplemental product reports, as required according to 21 CFR 1002.11.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Laser light show projectors manufactured by BLS, Model DIODE 1000/2000/3000.
- FDA recall number
- Z-2111-2025
- Quantity
- 10
Laser light show projectors manufactured by BLS, Model CLUB 1000/2000/4000.
- FDA recall number
- Z-2112-2025
- Quantity
- 12
Laser light show projectors manufactured by BLS, Model PURE 5000/10000.
- FDA recall number
- Z-2113-2025
- Quantity
- 3
Distribution (FDA): US Nationwide Distribution
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Beyond Laser Systems, LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.