Why it was recalled
Label is incorrect. The product label indicates that the device contains an air eliminating filter. However, the filter utilized on this extension set is not indicated for removal of air and does not feature an air vent, which is common to air eliminating filters.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
9 more codes are in the record below.
Microbore Extension Set with 0.2 Micron Filter. Catalog Number: V6215. for use with IV tubing.
- Lot or code
- Catalog Number: V6215. Primary UDI-DI 04046964189173, Unit of Use UDI-DI: 04046964189166. Lot Number (Expiration Date): 0061742452 (30JUN2025), 0061747379 (31AUG2025 0, 0061767411 (31JAN2026), 0061780914 (30APR2026), 0061803499 (30SEP2026), 0061806173 (30NOV2026), 0061822333 (30APR2027), 0061836578 (30APR2027), 0061849109 (31JUL2027), 0061850836 (31AUG2027), 0061899902 (30SEP2028), 0061936368 (31MAR2029), 0061936119 (31MAR2029), 0061936369 (31MAR2029), 0061940584 (30APR2029).
- FDA recall number
- Z-2176-2025
- Quantity
- 34,825 units US
Distribution (FDA): US distribution to AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, WA, WY. International distribution to Canada.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- B BRAUN MEDICAL INC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.