What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury of Remote probability. Only in some (very rare) cases may the issue occur.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
1 more codes are in the record below.
Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma.
- Lot or code
- Catalog Number DSL8700 UDI Code: 15099590211615 Serial/Lot Numbers: 241014D 241111C 241209C 241209D 250106C 250106D 250203C
- FDA recall number
- Z-2336-2025
- Quantity
- 279 units
Distribution (FDA): Worldwide - U.S. Nationwide distribution in the states of NC and the countries of Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, and Netherlands.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Immunotech A.S.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.