FDA recall 97075: SEDECAL SA: Sedecal Mobile Digital Diagnostic X-Ray Systems and 45 more

FDA recall 97075 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

391 more codes are in the record below.

  • Sedecal Mobile Digital Diagnostic X-Ray Systems; SM-40HF-B-D-KM (AeroDR X30)

    Lot or code
    UDI/DI 08436046002593, Serial Numbers: G-81478, G81924, G84618, G87981, G88006, G88007, G87866, G84607, G86618, G88041, G87427, G80991, G-81936, G86688, G-82904, G86360, G84340, G86372, G81471, G86358, G86361, G87844, G87845, G85932, G86683, G84620, G84621, G85914, G87409, G87434, G84326, G84325, G-81925, G81923, G-81935, G84619, G-81934, G87426, G81002, G86686, G84339, G88042, G-82885, G-82887, G-82905, G84327, G85788, G85913, G85801, G86356, G87867, G88008, G87425, G86659, G88009, G87846.
    FDA recall number
    Z-0242-2026
    Quantity
    56 units
  • MobileDiagnost wDR 2.2; Model Number REF 9890-010-89522, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002357, Serial Numbers: SN11000007, SN13000099, SN13000100, SN13000101, 13000079, SN13000080, SN13000082, SN13000062, SN13000059, SN13000063, SN13000061, 13000067, 13000076, SN13000089, SN13000103, 13000085, SN13000146, 13000094, 13000092, 13000093, SN13000087, SN13000096, SN13000095, SN13000098, SN13000097, SN13000159, SN13000148, 13000145, 13000117, 14000051, SN13000149, 13000168, 14000015, 14000017, 14000014, 14000013, 14000019, 14000016, 13000167, 13000169, 14000011, 12000166, SN13000170, 13000200, 14000083, 14000008, 14000009, 13000201, SN13000179, SN13000181, 14000052, 14000026, 61918783, 14000049, 11000010, 13000068, 14000047, 14000041, 62072683, 62150154, 14000020, 62415613, SN14000097, SN14000096, 62587177, 14000116, 14000109, 14000119, 14000118, 14000126, 63604709, SN14000128, 64123668, 64216085, 14100105, 14100104, SN15100010, SN15100013, 15100104, 64904292, 15100100, 15100042, 15100040, 15100041, 15100226, 15100095, 65073897, 15100198, 15100116, 15100115, 15100127, 15100131, 15100135, 15100138, 15100144, 15100133, 14100034, SN14000127, 15100191, 66690596, 66936848, 15100274, 15100280, 15100276, SN16100008, 16100011, 67905914, 16100037, 16100038, 68025890, 68052381, 68054442, 68199970, 68233849, 68268597, 12000085, 68530038, 16200084, 16200028, 16200029, 16200049, 16200050, 16200051, 69022933, 16200064, 16200099, 70240554, 70378102, 70544668, 70545417, 70705340, 71001729, 72299716, 72555189, 21410094, 21410261, 21410347, 21410348, 22410146, 97579477,…
    FDA recall number
    Z-0243-2026
    Quantity
    1198 units

    The text above is the beginning of the record; 12,530 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Model Number 40KWFX, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G71453, G70953, G70930, G70537, G70757, G70797, G70275, G71257, G71983, G71975, G71467, G69702, G71440, G71329, G70244, G70955, G71462, G71320, G71451, G71466, G71335, G71446, G65260, G70951, G70814, G70931, G70942, G71991, G70261, G72080, G69697, G71978, G70938, G70247, G70250, G70947, G69706, G78917, G70943, G71343, G71345, G71346, G71258, G70527, G69551, G71442, G71977, G71987, G72083, G70242, G72111, G70256, G70806, G71339, G69547.
    FDA recall number
    Z-0244-2026
    Quantity
    55 units
  • Model Number 40KW-FX.899, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G72770
    FDA recall number
    Z-0245-2026
    Quantity
    1 unit
  • Model Number 40KWFXPLUS, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G75176, G75504, G76697, G78126, G78025, G78715, G78307, G79071, G79091, G79193, G80165, G79928, G76507, G79861, G80255, G80202, G82594, G79057, G73017, G80154, G73616, G73471, G76479, G79615, G80179, G79467, G79827, G79840, G79877, G79926, G78011, G78726, G82574, G82638, G73651, G75438, G77820, G78050, G79426, G79634, G82590, G72392, G73520, G73524, G73627, G73642, G73653, G76730, G76888, G78212, G78279, G78296, G78305, G78323, G82156, G83117, G83133, G73624, G82557, G84603, G84633, G84634, G84641, G84642, G84753, G84778, G84795, G84857, G84869, G84882, G84884, G84900, G84901, G84902, G84911, G84912, G85513, G85514, G85516, G85527, G85528, G79115, G79915, G72362, G75112, G75515, G79299, G79324, G73453, G79625, G77328, G75676, G75744, G75439, G78129, G78141, G76110, G76545, G78000, G80042, G73460, G75100, G75502, G84782, G79794, G83972, G84025, G79936, G84487, G75404, G73457, G75085, G79927, G75388, G73609, G76560, G78992, G82079, G82095, G84014, G75375, G75503, G75516, G73621, G73648, G72971, G77213, G79889, G77605, G79311, G75961, G77789, G84028, G73015, G73063, G78873, G79663, G79275, G73027, G76887, G79277, G79357, G76126, G75980, G76125, G76844, G77073, G77089, G77191, G77425, G73079, G75867, G83920, G84017, G77609, G77653, G84591, G75904, G75905, G73147, G73059, G77583, G77641, G77768, G77854, G77870, G77928, G84038, G72363, G73052, G75661, G75662, G75691, G84052, G72972, G73084, G73119, G76757, G76803, G84066, G77780, G73040,…
    FDA recall number
    Z-0246-2026
    Quantity
    268 units

    The text above is the beginning of the record; 688 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Model Number 40KWFXPLUS.002, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G73508, G73512, G71323, G72419, G78910, G72425, G83214, G70934.
    FDA recall number
    Z-0247-2026
    Quantity
    8 units
  • Model Number 40KWFXPLUS.004, Mobile X-ray system

    Lot or code
    UDI/DI unknown, Serial Numbers: G84489, G84504, G84589.
    FDA recall number
    Z-0248-2026
    Quantity
    3 units
  • Model Number 40KWFXPLUS.005, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G82576, G82871, G72391, G74872, G75722, G76782, G75035, G75723, G78884, G70532, G73630, G82077.
    FDA recall number
    Z-0249-2026
    Quantity
    12 units
  • Model Number 40KWFXPLUS.889, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G77146
    FDA recall number
    Z-0250-2026
    Quantity
    1 unit
  • Model Number 40KWFXPLUS-710CW, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G83204, G82622, G79677, G79194, G82129, G79632, G79837, G79924, G84490, G84640, G79699, G84605.
    FDA recall number
    Z-0251-2026
    Quantity
    12 units
  • Model Number 40KWFXPLUS-710CW.007, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G79192, G79195, G79212, G78205, G80256, G79692, G80053.
    FDA recall number
    Z-0252-2026
    Quantity
    7 units
  • Model Number 40KWMOBILE3.899, Mobile X-ray system

    Lot or code
    UDI/DI unknown, Serial Numbers: G64852
    FDA recall number
    Z-0253-2026
    Quantity
    1 unit
  • Model Number SM-20HF-Batt; 20KW ANALOG, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001091, Serial Numbers: G23218
    FDA recall number
    Z-0254-2026
    Quantity
    1 unit
  • Model Number SM-32HF-B-D-C; 32KW 50G, Mobile X-ray system

    Lot or code
    UDI/DI unknown, Serial Numbers: G22915, G23633, G23828, G20560, G21319, G24800, G19667, G20347.
    FDA recall number
    Z-0255-2026
    Quantity
    8 units
  • Model SM-32HF-Batt; 32KW ANALOG, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001466, Serial Numbers: G-33344, G29404, G29407, G30258, G30236, G63124, G60169, G26234, G26284, G25734, G38069, G-40590, G41143, G62020, G69393, G38082, G41163, G60166, G26244, G40196, G60461, G33309.
    FDA recall number
    Z-0256-2026
    Quantity
    22 units
  • Model SM-40HF-B-D-C; 40KW - CANON READY, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G-33126, G31259, G35641, G35630, G35718, G37357.
    FDA recall number
    Z-0257-2026
    Quantity
    6 units
  • Model Number SM-40HF-B-D-C; 40KW 50G, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G25456, G25508, G25681, G25698, G25930, G27630, G27705, C-25709, G22652, G38643, G25673, G27664, G25708, G25689, G25457, C08042, G25641, G25663, G24234, G25517, G23833, G23945, G23954, G25458, G25528, G27497, G27497/584, G27651, G27689, G22586, G24220, G25755, G26016, G25529, G25396, G27792, C25709, G27701, G30684.
    FDA recall number
    Z-0258-2026
    Quantity
    39 units
  • Model Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G37528
    FDA recall number
    Z-0259-2026
    Quantity
    1 unit
  • Model SM-40HF-B-D-C; 40KW 70C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G13768, G36626, G62493, G39518, G61725, G36211, G36281, G36573, G36581, G61368, G61880, G62037, G61367, G36227.
    FDA recall number
    Z-0260-2026
    Quantity
    14 units
  • Model Number SM-40HF-Batt; 40KW ANALOG, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001114, Serial Numbers: G24948, G25512, C25673, 25901, G25771, G26990, G25506, G22189, G30240.
    FDA recall number
    Z-0261-2026
    Quantity
    9 units
  • Sedecal Easy Moving Plus Model Number SM-50HF-B-D-C; 50KW 50G, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001527, Serial Numbers: 151248, G23340, C05006-W, G22215, G27497/584, 7034, G25402, G25910.
    FDA recall number
    Z-0262-2026
    Quantity
    8 units
  • Model Number SM-40HF-B-D-C; 40KW 50C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G32487, 31414, G35462, G32657.
    FDA recall number
    Z-0263-2026
    Quantity
    4 units
  • Model Number SM-40HF-B-D-C; 40KW 50G, Mobile X-ray system

    Lot or code
    UDI/DI unknown, Serial Numbers:G31761, G30761, G32587, G32577, G31065, G33581, G30808, G30762, G29873, G30156, G30173, G30952, G30689, G33004, G30155, G30025, G35450, G35620, G30570, G31077, G29426, G33144, G29416, G30160, G30685, G33723, G30165.
    FDA recall number
    Z-0264-2026
    Quantity
    27 units
  • Model Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G35462, G-32753, G35480, G33111, G33721.
    FDA recall number
    Z-0265-2026
    Quantity
    5 units
  • Model Number SM-40HF-B-D-C; 40KW 55C/60C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G33585
    FDA recall number
    Z-0266-2026
    Quantity
    1 unit
  • Model Number SM-40HF-B-D-C; 40KW 55G, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G-33103, G35422, G35504, G33127, G33646.
    FDA recall number
    Z-0267-2026
    Quantity
    5 units
  • Model Number SM-40HF-B-D-C; 40KW 55G/60G, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G35524, G33711.
    FDA recall number
    Z-0268-2026
    Quantity
    2 units
  • Model Number SM-40HF-B-D-C; 40KW 55C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G37587.
    FDA recall number
    Z-0269-2026
    Quantity
    1 unit
  • Model Number SM-40HF-B-D-C; 40KW 55C/60C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G35611, G37536, G37588, G37490.
    FDA recall number
    Z-0270-2026
    Quantity
    4 units
  • Model Number SM-40HF-B-D-C; 40KW 70C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: D11-40KW-F, G36689, G37293, G38198, G38657, G38385, G37687, G36643, G41114, G36137, G36761, G36272, G37188, G37359, G37456, G38178, G62012, G62059, G62892, G62943, G41083, G41100, G36217, G36537, G60066, G61812, G61812, G63322, G40016, G41203, G41222, G60061, G60063, G60064, G36732, G36751, G41119, G62480, G60418, G61707, G65172, G39639, G41104, G36526, G36601, G37129, G37209, G41244, G62015, G40024, G62828, G60429, G61818, G61873, G60065, G61786, G37430, G36148, G39650, G64186, G65400, G64854, G62626, G62660, G62684, G62687, G62747, G62750, G62762, G62938, G63484, G62804, G64206, G64813, G64179, G62500, G62502, G62565, G62690, G62860, G62867, G62883, G62891, G62947, G36222, G60415, G60070, G37646, G61715, G39589, G37466, G38292, G38173, G38058, G38293, G60069, G62816, G63335, G36244, G36492, G37442, G38613, G40124, G40211, G38190, G40199, G62567, G39809, G38287, G37196, G41256, G61872, G62029, G64891, G62682, G62900, G61311, G37239, G38279, G37678, G38076, G61307, G39808, G40133, G62821, G64883, G60058, G60062, G61867, G37676, G40203, G37422, G62815, G36468, G37413, G40216, G41058, G41066, G63212, G41267, G40227, G60430, G63318, G63328, G63333, G41108, G41109, G61716, G60518, G60652, G61356, G60520, G60597, G39800, G39801, G60582, G39657, G62030, G65393, G65395, G39782, G39791, G61706, G61795, G62053, G62062, G62011, G64512, G64520, G65289, G60576, G60606, G60631, G61339, G39465, G61333, G61337, G38629, G62511, G41218, G60635.
    FDA recall number
    Z-0271-2026
    Quantity
    181 units
  • Model Number SM-40HF-B-D-C; 40KW 70C EN, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G62066
    FDA recall number
    Z-0272-2026
    Quantity
    1 unit
  • Model Number SM-40HF-B-D-C; 40KW 70C READY, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G37158, G61698, G62863, G41192, G62814, G41098, G62566, G60400, G63392, G60406.
    FDA recall number
    Z-0273-2026
    Quantity
    10 units
  • Model Number SM-40HF-B-D-C; 40KW 80C, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G39822, G62662, G62564, G61702, G60067, G38060.
    FDA recall number
    Z-0274-2026
    Quantity
    6 units
  • Model Number SM-40HF-B-D-C; 40KW CANON READY, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G41234
    FDA recall number
    Z-0275-2026
    Quantity
    1 unit
  • Model Number SM-40HF-B-D-C; 40KW CANON READY.005, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G36566
    FDA recall number
    Z-0276-2026
    Quantity
    1 unit
  • Model Number SM-40HF-B-D-C; 40KW, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G65281, G60407, G65682, G64582, G64789, G70752, G65384, G65418, G67250, G68470, G70521, G70524, G70913, G70914, G65166, G63507, G64205, G64842, G71447, G65175, G65382, G64793, G64798, G40226, G72434, G65554, G67230, G37280, G65174, G65558, 15033R, G64271, G64713, G64849, G65168, G71272, G71309, G65284, G72955, G65043, G70919, G72429, G72789, G70928, G72776, G72766, G63409, G63418, G64893, G66907, G65588, G65402, G73096, G73131, G71269, G71750, G71762, G71267, G71727, G71732, G71745, G71755, G71761, G66892, G65275, G70922, G73154, G64505, G65744, G73033, G73134, G64941, G63489, G65036, G64256, G72976, G73092, G73145, G64196, G64945, G72412, G73074, G65414, G71726, G71743, G64807, G66872, G67257, G40208, G64885, G38391, G39467, G39818, G65408, G70748, G37108, G6715, G67232, G71277, G71286, G64604, G37406, G40004, G63223, G63224, G63232, G61711, G71271, G71302, G72790, G64500, G65269, G65692, G65290, G65691, G66862, G65283, G64189, G70744, G64812, G70755, G65286, G65266, G72760, G66884, G64601, G71300, G67185, G63495, G64890, G72780, G64607, G63488, G63494, G71285 - NCR1161, G63501, G41245, G64797, G64819, G64855, G72803, G41213.
    FDA recall number
    Z-0277-2026
    Quantity
    142 units
  • Model Number SM-40HF-B-D-C; 40KW.002, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G39544, G62558, G71736, G40116, G65568, G65570.
    FDA recall number
    Z-0278-2026
    Quantity
    6 units
  • Model Number SM-40HF-B-D-C; 40KW.005, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G64274, G40012, G64597, G70742, G72764, G61804.
    FDA recall number
    Z-0279-2026
    Quantity
    6 units
  • Model Number SM-40HF-B-D-C; 40KW70CDEMO.005, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G38652, G37633, G36042, G37641, 608A, G36681.
    FDA recall number
    Z-0280-2026
    Quantity
    6 units
  • RADPRO MOBILE, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G30567, G31256, G32581, G30250, G28320, G30902, G29864, G35389, G30960, G33238, G31252, G33708, G30771, G33634, 31412, 31480, G33004, G33112, G33585, G31271, G31285, G32514, G32498, G35665.
    FDA recall number
    Z-0281-2026
    Quantity
    24 units
  • SED URS LP, Mobile X-ray system

    Lot or code
    UDI/DI unknown, Serial Numbers: 40116
    FDA recall number
    Z-0282-2026
  • SHIMADZU RF.004, Mobile X-ray system

    Lot or code
    UDI/DI unknown, Serial Numbers: 75723
    FDA recall number
    Z-0283-2026
  • SM_40HF_B_D_VIR3, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G86476, G86632, G86634, G85447, G86619, G85453, G85461, G86477, G86633, G87334, G87335, G87336, G87353, G86617, G87337, G85335, G85452.
    FDA recall number
    Z-0284-2026
    Quantity
    17 units
  • SM_40HF_B_D_VIR3.004, Mobile X-ray system

    Lot or code
    UDI/DI 08436046002166, Serial Numbers: G73023, G87352, G80066.
    FDA recall number
    Z-0285-2026
    Quantity
    3 units
  • SM_40HF_BDC, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: 31526
    FDA recall number
    Z-0286-2026
    Quantity
    1 unit
  • SM_40HF_BDC3, Mobile X-ray system

    Lot or code
    UDI/DI 08436046001510, Serial Numbers: G41213
    FDA recall number
    Z-0287-2026
    Quantity
    1 unit

Distribution (FDA): US: CA, IL, and NJ

Source

Issued by
U.S. Food and Drug Administration
Firm
SEDECAL SA
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: October 2025, Week 2025-W44. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97075 on October 29, 2025.