What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Glucose microcuvettes experienced transit time outside of limits, so stability through the whole microcuvette lifetime can no longer be guaranteed. The affected products may not have been stored at the recommended labeled storage conditions which may have impacted the safety, quality, identity, potency, and purity of the product. This failure could increase the probability of an incorrect test result or a delayed test result.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 (or DM) Analyzer with plasma conversion that multiplies the measured whole blood glucose value by a factor of 1.112 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. The box contains 4 vials with 25 microcuvettes in each vial.
- Lot or code
- UDI-DI: 07311091107060, Lot: 2501708 (Only 4 boxes within specific McKesson Shipment)
- FDA recall number
- Z-2070-2025
- Quantity
- 4 Boxes
Distribution (FDA): US Nationwide distribution in the state of AR.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- HemoCue AB
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.