Skip to content
FDAMedical devicesClass IOngoing

FDA recall 96943: Edwards Lifesciences, LLC: Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula… and 5 more

FDA recall 96943 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

180 more codes are in the record below.

  • Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No.

    Lot or code
    Lot Code: Model No:OPTI16; UDI: 00690103180558; All lots up to BSLC9149 Serial Numbers: 130467 130471 130602 130765 130816 130822 130852 130949 131039 131041 131044 131237 131312 131428 131509 131826 131887 131908 132338 132436 132503 132831 BSLC0040 BSLC0165 BSLC0165 BSLC0320 BSLC1309 BSLC1309 BSLC1468 BSLC1468 BSLC1557 BSLC1682 BSLC1688 BSLC1825 BSLC1830 BSLC5302 BSLC5899 BSLC6166 BSLC7142 BSLC7142 BSLC7513 BSLC8176 BSLC8659 BSLC9149
    FDA recall number
    Z-2080-2025
    Quantity
    27022
  • Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI18 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

    Lot or code
    Lot Code: Model No: OPTI18; UDI: 00690103180565; All lots up to: BSLC9155 Serial Numbers: 130451 130466 130472 130593 130709 130764 130818 130821 130847 130948 130952 131038 131043 131047 131182 131193 131231 131236 131289 131311 131317 131433 131435 131513 131516 131553 131554 131606 131623 131625 131628 131682 131737 131909 131929 131966 132175 132201 132366 132395 132416 132430 132435 132505 132832 132958 132959 BSLC0110 BSLC0166 BSLC0661 BSLC0661 BSLC0826 BSLC0826 BSLC0826 BSLC0826 BSLC0855 BSLC0855 BSLC0887 BSLC0887 BSLC0979 BSLC1114 BSLC1114 BSLC1211 BSLC1308 BSLC1469 BSLC1656 BSLC1660 BSLC1815 BSLC3221 BSLC4153 BSLC4400 BSLC4461 BSLC4513 BSLC5027 BSLC5027 BSLC5320 BSLC5367 BSLC5765 BSLC6421 BSLC6467 BSLC7335 BSLC7514 BSLC7515 BSLC8175 BSLC8385 BSLC8535 BSLC8657 BSLC9100 BSLC9104 BSLC9105 BSLC9152 BSLC9155
    FDA recall number
    Z-2081-2025
    Quantity
    13240
  • Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

    Lot or code
    Lot Code: Model No: FEMII016A; UDI: 00690103031232; All lots up to: BSLC8179 Serial Numbers: 130544 130757 131105 131313 131511 131535 131888 131925 132504 BSLC0184 BSLC0386 BSLC0596 BSLC1311 BSLC3741 BSLC4162 BSLC5040 BSLC6487 BSLC8179
    FDA recall number
    Z-2082-2025
    Quantity
    6801
  • Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

    Lot or code
    Lot Code: Model No; FEMII016AS; UDI: 0060913168341; All lots up to: BSLC1689 Serial Numbers: FEMII016AS FEMII016AS FEMII016AS
    FDA recall number
    Z-2083-2025
    Quantity
    847
  • Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

    Lot or code
    Lot Code: Model No: FEMII018A; UDI: 0069013031256; All lots up to: BSLC9148 Serial Numbers: 130468 130599 130704 130848 130953 131048 131286 131430 131626 131774 131775 132004 132005 132510 BSLC0025 BSLC0121 BSLC0422 BSLC1115 BSLC1470 BSLC4161 BSLC4472 BSLC5041 BSLC5293 BSLC5842 BSLC6163 BSLC6164 BSLC7336 BSLC7794 BSLC8660 BSLC9148
    FDA recall number
    Z-2084-2025
    Quantity
    13240
  • Brand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No

    Lot or code
    Lot Code: Model No: FEMII018AS; UDI: 00690103168358; All lots up to: BSLC9064 Serial Numbers: 131776 BSLC4190 BSLC9064
    FDA recall number
    Z-2085-2025
    Quantity
    558

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates (UAE), United Kingdom.

Source

Issued by
U.S. Food and Drug Administration
Firm
Edwards Lifesciences, LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Edwards Lifesciences, LLC, Month: July 2025, Week 2025-W29. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96943 on July 16, 2025.

Other notices of Edwards Lifesciences, LLC