What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent for reagents and human plasmas in coagulation tests performed with analyzers of the STA-R, STA Compact, STA Satellite and ST art families.
- Lot or code
- UDI: (01)13607450003601(11)240331(17)260331(10)270902(241)00360. Lot Number: 270902.
- FDA recall number
- Z-1993-2025
- Quantity
- 939 US ( 2,810 total worldwide)
Distribution (FDA): The affected lot was distributed to customers in various US states in October and November 2024. The lot was also distributed in other countries including Canada. The lot was provided to two Distributors in US Territories: MedPharm Corp (Guam) and Isla Lab Products (Puerto Rico); MedPharm will be notified by Stago's Asia-Pacific business unit in the coming days and is not part of this report.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Diagnostica Stago, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.