What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Contact lens insertion solution may lack sterility.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Contamac, Nutrifill Physiologic formula, Contact Lens Insertion solution
- Lot or code
- UDI: 00850012123002/ Lot: HFH, HGC, HGI, HGN, HKF, HKK
- FDA recall number
- Z-1956-2025
- Quantity
- 17,273 cartons (604,555 ampoules)
Distribution (FDA): US Nationwide distribution in the states of AZ, CA, CO, FL, IL, IN, KS, KY, LA, ME, MI, NC, NJ, NY, OH, OK, OR, PA, RI, SC, TX, UT, VA, WA, WI.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Contamac Solutions, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.