What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
221 more codes are in the record below.
Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system
- Lot or code
- REF/UDI-DI/Serial Number: K9303320/05415334002364/2531531598, 2531531604, 2531534397, 2531534420, 2531534430, 2531534466, 2531534905, 2531534948, 2531534956, 2531536564, 2531554560, 2531572616, 2531572628, 2531572634, 2531591962, 2531591973, 2531591992, 2531592064, 2531592096, 2531593814, 2531593818, 2531593823, 2531597597, 2531610546, 2531610547, 2531610572, 2531610578, 2531610580, 2531612404, 2531612420, 2531612424, 2531612430, 2531612431, 2531612432, 2531612433, 2531612434, 2531612436, 2531612442, 2531623109, 2531623114, 2531623131, 2531623133, 2531623141, 2531635578, 2531635582, 2531639224, 2531639261, 2531639263, 2531639353, 2531640594, 2531640597, 2531640598, 2531640603, 2531640626, 2531640634, 2531540640, 2531640646, 2531640658, 2531640660; K9303331/05415334026827/2531533275, 2531533289, 2531533292, 2531533293, 2531533304, 2531533310, 2531533312, 2531533314, 2531533316, 2531533353, 2531533356, 2531534284, 2531534292, 2531534362, 2531537718, 2531540799, 2531569738, 2531582287, 2531582290, 2531582311, 2531582315, 2531582378, 2531582750, 2531582755, 2531582790, 2531582791, 2531582826, 2531582827, 2531582829, 2531584018, 2531584044, 2531537696, 2531537726; K9303300/05415334002180/2531537815, 2531557845, 2531557866, 2531558588, 2531558727, 2531558738, 2531558753, 2531558754, 2531558759, 2531561761, 2531561785, 2531572541, 2531572557, 2531572568, 2531572576, 2531572847, 2531572862, 2531572872, 2531633606, 2531633662, 2530887517, 2530887515;…
- FDA recall number
- Z-1994-2025
- Quantity
- 223
The text above is the beginning of the record; 1,369 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): US Nationwide distribution in the states of TX, OH. NJ.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Barco N.V.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.