What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Completed. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3
- Lot or code
- UDI-DI (01)00850038042028; Lot Numbers: 1020040211, 1020040212, 1020040213
- FDA recall number
- Z-1905-2025
- Quantity
- 3 units
Distribution (FDA): US Distribution to States: CA, FL, NE
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- VIRTUAL INCISION CORPORATION
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.