Why it was recalled
HKH 8820 Wall Holder does not comply with standard DIN EN 1789:20 and has the potential to detach.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Cardiohelp System HKH 8820 Wall Holder. Mounting bracket designed to hold the CARDIOHELP-i system for interhospital transport by vehicle.
- Lot or code
- Model No. 701045366; UDI 04037691456584; All Serial No.
- FDA recall number
- Z-1823-2025
- Quantity
- US: 2 units; OUS: 168 units
Distribution (FDA): Domestic: IL & MO. Foreign: Australia, Austria, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, South Korea, Poland, Russia, Saudi Arabia, Spain, Switzerland, Thailand.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Maquet Cardiopulmonary Ag
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.