What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
ELEOS COLLAR STEM, CEMENTED, FLUTED, 13MM X 120MM. Model Number: HC-13120-03M. Component of Limb Salvage System with BioGrip.
- Lot or code
- Model Number: HC-13120-03M. UDI-DI: B278HC1312003M0. Serial Numbers: P230274-1XX (XX represents 23 parts in sequential order from 101-123).
- FDA recall number
- Z-1817-2025
- Quantity
- 23 units
ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip.
- Lot or code
- Model Number: HC-15120-03M. UDI-DI: B278HC1512003M0. Serial Numbers: P230274-1XX (XX represents 23 parts in sequential order from 101-123).
- FDA recall number
- Z-1818-2025
- Quantity
- 23 units
Distribution (FDA): US Nationwide distribution in the states of California, Kentucky, and Wisconsin.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Onkos Surgical, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.