What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Thrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Artix MT Thrombectomy Device, REF: 32-102
- Lot or code
- UDI-DI: 00810123710584, Lot: 24100082, Expiration: 14 April 2025. IFU: IU-01087 Rev. B
- FDA recall number
- Z-1728-2025
- Quantity
- 45
Distribution (FDA): US Nationwide distribution in the states of FL, IL, WI, KY, TX, CA, OH, GA, AK, MI, PA, WA, CO, OH.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Inari Medical - Oak Canyon
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.