What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
- Lot or code
- UDI/DI 10841736116071, All Serial Numbers
- FDA recall number
- Z-1663-2025
- Quantity
- 112 units
Distribution (FDA): US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- IHB OPERATIONS B.V.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.