What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK000196, 9174118, CK000955, 3174135, CK000617, CK000839, 3194355, 6174118, 3194135, 3194335, 2174108, 9194118, 3194118, 1174118, CK000285C, CK000335A, CK000796, 1194108D, CK000757, CK001042, 9174108, 1174108D1, CK000631B, 3194155, 3174108, CK000567, 1194108D4, 1174108D2, CK000677, CK000864, CK000937A, CK000930, CK001027, 1174108D, CK000867, CK000900, CK000315A, CK000589A, CK000854B, CK000897, CK000953A, CK000586B, CK000549, CK000844, 1194108D1, 1174108D4, 1194108D3, CK000934, CK000986, CK000411, CK001018, 1174108, CK000060, CK000513A, CK001021, CK001030, CK000501A, CK000619A, CK000903, 3194108, 1194108D2, 9194108D, CK000959, CK000613, CK000991, CK000983, CK000881, CK000721, CK000642, CK000289C, CK000894A, CK000914, CK000445, CK000687, CK000524A, CK000819, CK000266A, 1174108D3, 3194108D, CK000545, 1174108D5, 3174108D, CK000775A, 1194108D5, CK000736, CK000739, 9194108, 1194108, CK000660, CK000609A, 9174108D, CK001067, CK000077D, CK000882, CK000874, CK000994, CK000870A, CK000443C, CK000668B, CK000784
- Lot or code
- REF/UDI-DI/Lot(Expiration): 3174155/00801741027611/REGQ0844(6/30/2025), REGZ1439(11/30/2025), REHP0860(11/30/2025), REHQ3097(6/30/2026), REHR0022(6/30/2026), REHT1151(7/31/2026), REHU1525(11/30/2026), REHU2455(6/30/2026), REHU2614(11/30/2026), REHV2765 (7/31/2026), REJN3472(2/28/2027), REJP4216(5/31/2027), REJQ2272(7/31/2027), REJR1411(8/31/2027), REJR2565(4/30/2027), REJR2614(8/31/2027), REJT0755(9/30/2027); CK000540/00801741105142/REGZ1425(10/31/2025), REHP0808(11/30/2025), REHU1513(1/31/2026), REHV2221(1/31/2026), REHY2842(10/31/2026), REJN4145 (12/31/2026), REJR1322(3/31/2027), REJS2110(4/30/2027), REJS3773(2/28/2027); 3174118/00801741027598/REGZ2512(11/30/2025), REHP1760(1/31/2026), REHP2603(1/31/2026), REJN1059(12/31/2026); CK000483/00801741099397/REHN3187(10/31/2025), REHQ1804(6/30/2026), REJN1065(5/31/2027), REJP2018(6/30/2027), REJR1321(8/31/2027); 3174335/00801741027628/REHN3908(10/31/2025), REHP0800(10/31/2025), REHQ3070(11/30/2025), REHT1129(5/31/2026), REHU1519(6/30/2026), REHV1391(7/31/2026), REHV3050(7/31/2026), REHZ1011(11/30/2026), REJN4109(12/31/2026), REJQ0405(2/28/2027), REJR2090(3/31/2027), REJT0833(5/31/2027); 3174355/00801741027635/REHQ1781(11/30/2025), REHS1925(4/30/2026), REHT1120(5/31/2026), REHU0975(6/30/2026), REHU2488(6/30/2026), REHV1400(7/31/2026), REHV3020(7/31/2026), REHW1329(8/31/2026), REHY1684(10/31/2026), REHZ1903(11/30/2026), REJN3007(12/31/2026), REJP0855(1/31/2027), REJP2021(1/31/2027), REJQ2284(2/28/2027), REJR2157(3/31/2027),…
- FDA recall number
- Z-1507-2025
- Quantity
- 718,461
The text above is the beginning of the record; 10,299 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide - US Nationwide distribution including in the states of WI, NC, TX, NY, CA, MN, MT, RI, MA, OH, CT, SC, IL, CO, VA, NJ, GA, UT, FL, PA, IN, MD, KY, AZ, MI, WA, HI, MO, VT, AL, NM, TN, DE, LA, IA, NH, ID, DC, WV, KS, NE, SD, ND, WY, AK, OK, OR, AR and the countries of Andorra, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, French Polynesia, French Guayana, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Martinique, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Poland, Portugal, Qatar, Republic of Korea, Reunion, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Bard Access Systems, Inc.