What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The RSV target may give a late Ct value and could potentially not pass QC.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Helix Elite Inactivated Standard: Inactivated Influenza A/B and Respiratory Syncytial Virus, Multi-analyte, unassayed quality control material, Catalog Number HE0044N
- Lot or code
- UDI-DI: 10845357042184; Lot Number: HE0044-131
- FDA recall number
- Z-1454-2025
- Quantity
- 28 units
Distribution (FDA): Worldwide - US Nationwide distribution in the states of KY, MT, NJ, NY, WI and the countries of Ireland, Germany, UK.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Microbiologics Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.