What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
MIM software; System, Image Processing, Radiological
- Lot or code
- Software versions: 1. MIM Anyware 1.1, UDI/DI Not Applicable 2. MIM 7.2.0, UDI/DI 00850009343192; 3. MIM 7.2.1, UDI/DI 00850009343192; 4. MIM 7.2.2, UDI/DI 00850009343192; 5. MIM 7.2.3, UDI/DI 00850009343192; 6. MIM 7.2.4, UDI/DI 00850009343192; 7. MIM 7.2.5, UDI/DI 00850009343192; 8. MIM 7.2.6, UDI/DI 00850009343192
- FDA recall number
- Z-1267-2025
- Quantity
- 352 units
Distribution (FDA): Worldwide distribution - United States Nationwide and the countries of Australia, Belgium, Bolivia, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Israel, Italy, Japan, Mexico, Netherlands, Poland, Russia, Spain, Sweden, United Kingdom, Finland, Hong Kong.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- MIM Software Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.