What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Product may contain surface and subsurface contamination of Listeria monocytogenes which is typically visible immediately upon removal from the packaging, appearing on the blood side of the biplate.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
remel BLOOD AGAR MacCONKEY AGAR BIPLATE, REF R02049, QTY 100, for invitro diagnostic use
- Lot or code
- UDI/DI 848838003691, Lot Number 213926, exp. 2025-02-24
- FDA recall number
- Z-1200-2025
- Quantity
- 787 units
remel BLOOD AGAR EMB AGAR BIPLATE, REF R02041, QTY 100, for invitro diagnostic use
- Lot or code
- UDI/DI 848838003684, Lot Number 213971, exp. 2025-02-25
- FDA recall number
- Z-1201-2025
- Quantity
- 316 units
Distribution (FDA): US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Remel, Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.