Why it was recalled
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
- Lot or code
- Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO; UDI-DI: (1) no UDI, (2) 00812394021154, (3) no UDI, (4) no UDI, (5) 00812394022243, (6) 00812394021192, (7) 00812394021390; Serial Numbers: (1) D00000177884, D00000177917, D00000177931, D00000177956, D00000178271, D00000178311, D00000178395, D00000178542, M00000002041, M00000008225, M00000008290, M00000008336, (2) D00000003482, D00000003546, D00000003557, D00000003600, D00000003625, D00000003783, D00000003978, D00000003983, D00000003985, D00000004036, D00000004037, D00000004040, D00000004043, D00000004089, D00000004098, D00000004100, D00000004105, D00000004107, D00000004109, D00000004111, D00000004116, D00000004118, D00000004119, D00000004120, D00000004123, D00000004129, D00000004131, D00000004132, D00000004133, D00000004138, D00000004140, D00000004141, D00000004142, D00000004143, D00000004145, D00000004146, D00000004147, D00000004148, D00000004150, D00000004155, D00000004157, D00000004159, D00000004160, D00000004161, D00000004162, D00000004163, D00000004164, D00000004165, D00000004166, D00000004167, D00000004168, D00000004169, D00000004170, D00000004171, D00000004172, D00000004173, D00000004174, D00000004175, D00000004176, D00000004177, D00000004178, D00000004179, D00000004180, D00000004181, D00000004182, D00000004183, D00000004184, D00000004185, D00000004186, D00000004187, D00000004188, D00000004189, D00000004190, D00000004191, D00000004192, D00000004193,…
- FDA recall number
- Z-1316-2025
- Quantity
- 2268 units
The text above is the beginning of the record; 30,532 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
- Lot or code
- Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO; UDI-DI: (1) 00812394021192, (2) 00812394021390; Serial Numbers: D00000003715, D00000003785, D00000003909, D00000003916, D00000003920, D00000003937, D00000003940, D00000003946, D00000003948, D00000003953, D00000003969, D00000003975, D00000004007, D00000004008, D00000004009, D00000004010, D00000004012, D00000004014, D00000004016, D00000004017, D00000004018, D00000004020, D00000004023, D00000004024, D00000004025, D00000004027, D00000004028, D00000004029, D00000004030, D00000004032, D00000004033, D00000004034, D00000004035, D00000004038, D00000004039, D00000004041, D00000004042, D00000004044, D00000004045, D00000004047, D00000004049, D00000004050, D00000004051, D00000004052, D00000004053, D00000004054, D00000004055, D00000004056, D00000004057, D00000004079, D00000004097, D00000004103, D00000004106, D00000004108, D00000004122, D00000004124, D00000004127, D00000004134, D00000004135, D00000004136, D00000004137, D00000004139, D00000004144, D00000004195, D00000004198, D00000004200, D00000004202, D00000004210, D00000004212, D00000004218, D00000004222, D00000004223, D00000004228, D00000004232, D00000004236, D00000004237, D00000004239, D00000004244, D00000004245, D00000004250, D00000004251, D00000004252, D00000004256, D00000004260, D00000004264, D00000004265, D00000004275, D00000004307, D00000004315, D00000004377, D00000004380, D00000004382, D00000004383, D00000004385, D00000004386, D00000004387, D00000004391, D00000004392,…
- FDA recall number
- Z-1317-2025
- Quantity
- 56,981 units
The text above is the beginning of the record; 796,387 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80C, (2) G5A-80C-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
- Lot or code
- Model/Catalog Number: (1) G5A-80C, (2) G5A-80C-TSO; UDI-DI: (1) 00812394021208, (2) 00812394021406; Serial Numbers: (1) D00000022708, D00000033165, D00000033168, D00000033171, D00000033174, D00000033185, D00000033187, D00000033189, D00000033191, D00000033208, D00000033209, D00000033211, D00000033213, D00000033214, D00000033216, D00000033217, D00000033219, D00000033220, D00000033221, D00000033222, D00000033230, D00000035201, D00000035208, D00000035246, D00000035445, D00000035512, D00000035517, D00000035521, D00000035523, D00000035524, D00000035525, D00000035528, D00000035530, D00000035532, D00000035534, D00000035536, D00000035538, D00000035539, D00000035542, D00000035544, D00000035546, D00000035548, D00000035549, D00000035554, D00000035558, D00000035560, D00000035562, D00000035584, D00000035585, D00000035588, D00000035589, D00000035591, D00000035594, D00000035595, D00000035599, D00000035601, D00000035603, D00000035605, D00000035607, D00000035610, D00000035614, D00000035825, D00000035839, D00000035877, D00000036255, D00000036259, D00000036262, D00000036264, D00000036267, D00000036274, D00000036309, D00000036324, D00000036334, D00000036337, D00000036339, D00000036360, D00000036472, D00000036509, D00000036640, D00000036729, D00000036731, D00000036732, D00000036746, D00000036748, D00000036749, D00000036750, D00000036751, D00000036752, D00000036754, D00000036755, D00000036756, D00000036757, D00000036758, D00000036759, D00000036874, D00000036875, D00000036877, D00000036878,…
- FDA recall number
- Z-1318-2025
- Quantity
- 63,607 units
The text above is the beginning of the record; 889,153 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-01C, (2) G5A-02C, (3) G5A-05C, (4) G5A-08C, (5) G5A-10C, (6) G5A-13C, (7) G5A-15C, (8) G5A-22C, (9) G5A-41C, (10) G5A-85C, (11) G5A-90C (12) G5A-80-L; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
- Lot or code
- Model/Catalog Number: (1) G5A-01C, (2) G5A-02C, (3) G5A-05C, (4) G5A-08C, (5) G5A-10C, (6) G5A-13C, (7) G5A-15C, (8) G5A-22C, (9) G5A-41C, (10) G5A-85C, (11) G5A-90C (12) G5A-80-L; UDI-DI: (1) 00812394021161, (2) no UDI, (3) no UDI, (4) no UDI, (5) no UDI, (6) no UDI, (7) no UDI, (8) no UDI, (9) no UDI, (10) 00812394021208 (11) no UDI, (12) 00812394021215; Serial Numbers: (1) D00000066618, D00000066630, D00000066659, D00000066773, D00000067078, D00000067079, D00000067081, D00000067082, D00000067086, D00000067088, D00000067092, D00000067094, D00000067097, D00000067099, D00000067100, D00000067103, D00000067108, D00000067109, D00000067114, D00000067117, D00000067129, D00000067133, D00000067135, D00000067136, D00000067138, D00000067139, D00000067143, D00000067145, D00000067171, D00000067172, D00000067174, D00000067178, D00000067180, D00000067183, D00000067184, D00000067186, D00000067192, D00000067195, D00000067197, D00000067199, D00000067204, D00000067209, D00000067212, D00000067217, D00000067219, D00000067224, D00000067233, D00000067235, D00000067267, D00000067303, D00000067384, D00000067456, D00000067577, D00000067601, D00000067614, D00000067638, D00000067640, D00000067644, D00000067646, D00000067651, D00000067652, D00000067653, D00000067654, D00000067655, D00000067658, D00000067660, D00000067664, D00000067665, D00000067666, D00000067667, D00000067668, D00000067669, D00000067696, D00000067697, D00000067714, D00000067748, D00000067755, D00000067756, D00000067758, D00000067762,…
- FDA recall number
- Z-1319-2025
- Quantity
- 960 units
The text above is the beginning of the record; 12,367 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-01C, (2) G5S-02C, (3) G5S-10C, (4) G5S-11C; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
- Lot or code
- Model/Catalog Number: (1) G5S-01C, (2) G5S-02C, (3) G5S-10C, (4) G5S-11C; UDI-DI: (1) 00812394021246, (2) no UDI, (3) no UDI, (4) no UDI; Serial Numbers: (1) D00000089561, D00000089566, D00000089577, D00000105587, D00000111542, D00000111556, D00000111640, D00000111706, D00000111718, D00000113136, D00000113147, D00000113175, D00000113177, D00000113178, D00000113180, D00000113181, D00000113182, D00000113199, D00000113200, D00000113201, D00000113205, D00000113206, D00000113207, D00000134850, D00000134892, D00000135125, D00000136870, D00000136917, D00000136954, D00000157288, D00000170036, D00000170073, D00000172554, D00000178309, D00000178332, D00000183632, D00000191330, D00000191339, D00000191347, D00000191349, D00000191362, D00000191363, D00000191366, D00000221153, D00000221179, D00000221199, D00000221200, D00000221206, D00000221208, D00000221211, D00000221235, D00000221237, D00000221245, D00000221247, D00000221257, D00000221263, D00000221265, D00000221267, D00000221273, D00000221279, D00000221283, (2)D00000004566, D00000004687, D00000004688, D00000005214, D00000017929, D00000067528, D00000067585, D00000067595, D00000067685, D00000155123, (3) D00000171086, (4) M00000005364;
- FDA recall number
- Z-1320-2025
- Quantity
- 73 units
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80C; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
- Lot or code
- Model/Catalog Number: G5S-80C; UDI-DI: 00812394021284; Serial Numbers: D00000171086, D00000171087, D00000171088, D00000171089, D00000171090, D00000171092, D00000171093, D00000171094, D00000171095, D00000171096, D00000171097, D00000171098, D00000171099, D00000171100, D00000171101, D00000171102, D00000171103, D00000171105, D00000171107, D00000171109, D00000171111, D00000171113, D00000171115, D00000171117, D00000171118, D00000171121, D00000171126, D00000171133, D00000171134, D00000171136, D00000171138, D00000171140, D00000171144, D00000171147, D00000171150, D00000171154, D00000171157, D00000171159, D00000171160, D00000171164, D00000171165, D00000171166, D00000171167, D00000171168, D00000171169, D00000171170, D00000171171, D00000171172, D00000171178, D00000171179, D00000171181, D00000171183, D00000171186, D00000171188, D00000171190, D00000171601, D00000172373, D00000172376, D00000172379, D00000172381, D00000172382, D00000172383, D00000172384, D00000172385, D00000172386, D00000172391, D00000172412, D00000172413, D00000172414, D00000172416, D00000172421, D00000172422, D00000172425, D00000172426, D00000172428, D00000172430, D00000172431, D00000172432, D00000172433, D00000172434, D00000172435, D00000172436, D00000172437, D00000172438, D00000172439, D00000172444, D00000172446, D00000172448, D00000172449, D00000172455, D00000172466, D00000172473, D00000172482, D00000172483, D00000172484, D00000172486, D00000172487, D00000172488, D00000172489, D00000172493, D00000172497, D00000172503,…
- FDA recall number
- Z-1321-2025
- Quantity
- 7,221 units
The text above is the beginning of the record; 99,669 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-00A-TSO, (2) G5S-01A, (3) G5S-02A, (4) G5S-05A, (5) G5S-08A, (6) G5S-10A, (7) G5S-11A, (8) G5S-80A-TSO, (9) G5S-80-L (10) G5S-90A; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
- Lot or code
- Model/Catalog Number: (1) G5S-00A-TSO, (2) G5S-01A, (3) G5S-02A, (4) G5S-05A, (5) G5S-08A, (6) G5S-10A, (7) G5S-11A, (8) G5S-80A-TSO, (9)G5S-80-L, (10) G5S-90A; UDI-DI: (1) no UDI, (2) 00812394021239, (3) no UDI, (4) no UDI, (5) no UDI, (6) no UDI, (7) no UDI, (8) 00812394021413, (9) no UDI, (10) no UDI; Serial Numbers: (1) D00000154575, D00000154617, (2) D00000003725, D00000003727, D00000003734, D00000003896, D00000003898, D00000003923, D00000003927, D00000003941, D00000003943, D00000003951, D00000003955, D00000003962, D00000003964, D00000003965, D00000003968, D00000003971, D00000003972, D00000003981, D00000003982, D00000003984, D00000003987, D00000003989, D00000003991, D00000003997, D00000003999, D00000004149, D00000004151, D00000004153, D00000004154, D00000004156, D00000004158, D00000004301, D00000004304, D00000004308, D00000004313, D00000004324, D00000004329, D00000004365, D00000004390, D00000004394, D00000004402, D00000004407, D00000004408, D00000004410, D00000004413, D00000004427, D00000004441, D00000004464, D00000004509, D00000004515, D00000004516, D00000004654, D00000004658, D00000004659, D00000004669, D00000004672, D00000004673, D00000004698, D00000004703, D00000004716, D00000004841, D00000004855, D00000004857, D00000004869, D00000004880, D00000004889, D00000004891, D00000004897, D00000005194, D00000005204, D00000005207, D00000005208, D00000005218, D00000005220, D00000005230, D00000005235, D00000007936, D00000007937, D00000007938, D00000007942, D00000007961,…
- FDA recall number
- Z-1322-2025
- Quantity
- 1312 units
The text above is the beginning of the record; 17,239 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80A; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). There are two models available fully automatic and semi-automatic. After the defibrillations pads are applied to the patient, the fully automatic model evaluates the heart rhythm and, if a shockable rhythm is detected, delivers a shock without any rescuer assistance. The semi-automatic model evaluates the heart rhythm and requires the rescuer to press the shock button if a shockable rhythm is detected. Both models have voice and text instructions that guide the rescuer through the entire defibrillation process. Note: Not all configurations described in this document are available in all areas.
- Lot or code
- Model/Catalog Number: G5S-80A; UDI-DI: 00812394021277; Serial Numbers: D00000003974, D00000004048, D00000004128, D00000004404, D00000004436, D00000004556, D00000004560, D00000004563, D00000004564, D00000004573, D00000004581, D00000004582, D00000004642, D00000004665, D00000004666, D00000004718, D00000004730, D00000004736, D00000004743, D00000004744, D00000004750, D00000004770, D00000004774, D00000004779, D00000004780, D00000004784, D00000004785, D00000004787, D00000004789, D00000004794, D00000004799, D00000004800, D00000004801, D00000004802, D00000004805, D00000004806, D00000004807, D00000004808, D00000004809, D00000004810, D00000004812, D00000004815, D00000004816, D00000004818, D00000004820, D00000004821, D00000004824, D00000004873, D00000004916, D00000004929, D00000004933, D00000004939, D00000004940, D00000004951, D00000004955, D00000004960, D00000004970, D00000004974, D00000004976, D00000004983, D00000006293, D00000006294, D00000006296, D00000006338, D00000006341, D00000006343, D00000006851, D00000006852, D00000006893, D00000006894, D00000006904, D00000006915, D00000006918, D00000006921, D00000006925, D00000006928, D00000006931, D00000006975, D00000006976, D00000006989, D00000006990, D00000006991, D00000006992, D00000006994, D00000006998, D00000007012, D00000007237, D00000007238, D00000007239, D00000007241, D00000007242, D00000007243, D00000007244, D00000007245, D00000007249, D00000007252, D00000007254, D00000007255, D00000007256, D00000007257, D00000007258, D00000007259,…
- FDA recall number
- Z-1323-2025
- Quantity
- 8,744 units
The text above is the beginning of the record; 120,991 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Domestic: Nationwide Distribution International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA, PE, PG, PH, PK, PL, PR, PT, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, SM, TC, TH, TN, TR, TW, UA, US, VG, ZA.