What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
2 more codes are in the record below.
(1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser System.
- Lot or code
- Software version 2.4.1.0. (1) REF GA-2009996 - Serial number 1791, UDI-DI 07290109145518; (2) REF GA-0006802 - Serial number 1803, UDI-DI 07290109145525; and (3) REF-GA0008700 - Serial number 1818, UDI-DI 07290109145464.
- FDA recall number
- Z-1290-2025
- Quantity
- 3 devices
Distribution (FDA): Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- LUMENIS, LTD.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.