What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Error in the IFU lists an incorrect incubation time on one occasion of 20 minutes rather than the correct incubation time of 60 minutes.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
1 more codes are in the record below.
ASP BIOTRACE Auto Read 60 Steam BI Process Challenge Device, REF 73135.
- Lot or code
- Lot numbers: F40062, F40064, F40072, F40074, F40101, F40116; UDI-DI - 07798375772624 (primary packaging), 07798375772877 (secondary packaging) containing IFU Rev.A R0.
- FDA recall number
- Z-1248-2025
- Quantity
- 914 devices in the U.S.
Distribution (FDA): US Nationwide distribution in the states of CA and KY.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- TERRAGENE S.A.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.