Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 96120: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Allura CV20 System Code: 722031 and 18 more

FDA recall 96120 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

  • Allura CV20 System Code: 722031

    Lot or code
    UDI: N/A
    FDA recall number
    Z-1076-2025
    Quantity
    119 units OUS
  • Allura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722026

    Lot or code
    UDI: (1) 722003 00884838059191(10); (2) 722010 00884838059030(21); (3)722026 00884838054189(21).
    FDA recall number
    Z-1077-2025
    Quantity
    474 units US; 1908 units OUS
  • AlluraXper FD10/10 System Code: (1) 722005 (2) 722011 (3) 722027

    Lot or code
    UDI: (1) 722005 00884838059191(10) 00884838080805(10) (2) 722011 --- (3) 722027 00884838054196(21)
    FDA recall number
    Z-1078-2025
    Quantity
    146 units US; 629 units OUS
  • Allura Xper FD10C System Code: (1) 722001

    Lot or code
    UDI: N/A
    FDA recall number
    Z-1079-2025
    Quantity
    1 unit OUS
  • AlluraXper FD20 System Code: (1) 722006 (2) 722012 (3) 722028

    Lot or code
    UDI: (1) 722006 00884838055513(21) 00884838059115(21) 00884838059191(10) (2) 722012 00884838059054(21) (3) 722028 00884838054202(21)
    FDA recall number
    Z-1080-2025
    Quantity
    1528 units US; 3482 units OUS
  • AlluraXper FD20 Biplane System Code: (1) 722008 (2) 722013

    Lot or code
    UDI: (1) 722006 00884838055513(21) 00884838059115(21) 00884838059191(10) (2) 722012 00884838059054(21) (3) 722028 00884838054202(21)
    FDA recall number
    Z-1081-2025
    Quantity
    66 units US; 79 units OUS
  • AlluraXper FD20 Biplane OR Table System Code: (1) 722020 (2) 722025

    Lot or code
    UDI: N/A
    FDA recall number
    Z-1082-2025
    Quantity
    1 unit OUS
  • AlluraXper FD20 OR Table System Code: (1) 722023 (2) 722035

    Lot or code
    UDI: N/A
    FDA recall number
    Z-1083-2025
    Quantity
    1 unit US; 1 unit OUS
  • AlluraXperFD20/10 System Code: (1) 722029

    Lot or code
    UDI: 00884838054219(21)
    FDA recall number
    Z-1084-2025
    Quantity
    22 units US; 65 units OUS
  • Allura Xper FD20/20 System Code: (1) 722038

    Lot or code
    UDI: 00884838054226(21)
    FDA recall number
    Z-1085-2025
    Quantity
    33 units US; 82 units OUS
  • AlluraXperFD20/15 System Code: (1) 722058

    Lot or code
    UDI: 00884838059115(21)
    FDA recall number
    Z-1086-2025
    Quantity
    69 units US; 235 units OUS
  • Azurion 3 M12 System Code: (1) 722063 (2) 722221

    Lot or code
    UDI: (1) 722063 00884838085275(21) (2) 722221 00884838099203(21)
    FDA recall number
    Z-1087-2025
    Quantity
    9 units US; 101 units OUS
  • Azurion 3 M15 System Code: (1) 722064 (2) 722222 (3) 722280

    Lot or code
    UDI: (1) 722064 00884838085282(21); (2) 722222 00884838099210(21); (3) 722280 00884838103276(21)
    FDA recall number
    Z-1088-2025
    Quantity
    42 units US; 661 units OUS
  • Azurion 5 M12 System Code: (1) 722227 (2) 722231

    Lot or code
    UDI: (1) 722227 00884838099227(21); (2) 722231 00884838116740(21)
    FDA recall number
    Z-1089-2025
    Quantity
    6 units US; 246 units OUS
  • Azurion 5 M20 System Code: (1) 722228 (2) 722232 (3) 722281

    Lot or code
    UDI: (1) 722228 00884838099234(21); (2) 722232 00884838116757(21); (3) 722281 00884838110564(21)
    FDA recall number
    Z-1090-2025
    Quantity
    68 units US; 438 units OUS
  • Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235

    Lot or code
    UDI: (1) 722067 00884838085350(21); (2) 722225 00884838099265(21); (3) 722235 00884838116788(21)
    FDA recall number
    Z-1091-2025
    Quantity
    90 units US; 458 units OUS
  • Azurion 7 B20 System Code: (1) 722068 (2) 722226 (3) 722236

    Lot or code
    UDI: (1) 722068 00884838085367(21); (2) 722226 00884838099272(21); (3) 722236 00884838116801(21)
    FDA recall number
    Z-1092-2025
    Quantity
    251 units US; 938 units OUS
  • Azurion 7 M12 System Code: (1) 722078 (2) 722223 (3) 722233

    Lot or code
    UDI: (1) 722078 00884838085251(21); (2) 722223 00884838099241(21); (3) 722233 00884838116764(21)
    FDA recall number
    Z-1093-2025
    Quantity
    324 units US; 1390 units OUS
  • Azurion 7 M20 System code: (1) 722224 (2) 722079 (3) 722234 (4) 722282 System Code: (1) 722078 (2) 722223 (3) 722233

    Lot or code
    UDI: (1) 722224 00884838099258(21); (2) 722079 00884838085268(21); (3) 722234 00884838116771(21); (4) 722282 00884838110571(21)
    FDA recall number
    Z-1094-2025
    Quantity
    1346 units US; 2582 units

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.

Source

Issued by
U.S. Food and Drug Administration
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Month: February 2025, Week 2025-W07. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96120 on February 12, 2025.