What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Allura CV20 System Code: 722031
- Lot or code
- UDI: N/A
- FDA recall number
- Z-1076-2025
- Quantity
- 119 units OUS
Allura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722026
- Lot or code
- UDI: (1) 722003 00884838059191(10); (2) 722010 00884838059030(21); (3)722026 00884838054189(21).
- FDA recall number
- Z-1077-2025
- Quantity
- 474 units US; 1908 units OUS
AlluraXper FD10/10 System Code: (1) 722005 (2) 722011 (3) 722027
- Lot or code
- UDI: (1) 722005 00884838059191(10) 00884838080805(10) (2) 722011 --- (3) 722027 00884838054196(21)
- FDA recall number
- Z-1078-2025
- Quantity
- 146 units US; 629 units OUS
Allura Xper FD10C System Code: (1) 722001
- Lot or code
- UDI: N/A
- FDA recall number
- Z-1079-2025
- Quantity
- 1 unit OUS
AlluraXper FD20 System Code: (1) 722006 (2) 722012 (3) 722028
- Lot or code
- UDI: (1) 722006 00884838055513(21) 00884838059115(21) 00884838059191(10) (2) 722012 00884838059054(21) (3) 722028 00884838054202(21)
- FDA recall number
- Z-1080-2025
- Quantity
- 1528 units US; 3482 units OUS
AlluraXper FD20 Biplane System Code: (1) 722008 (2) 722013
- Lot or code
- UDI: (1) 722006 00884838055513(21) 00884838059115(21) 00884838059191(10) (2) 722012 00884838059054(21) (3) 722028 00884838054202(21)
- FDA recall number
- Z-1081-2025
- Quantity
- 66 units US; 79 units OUS
AlluraXper FD20 Biplane OR Table System Code: (1) 722020 (2) 722025
- Lot or code
- UDI: N/A
- FDA recall number
- Z-1082-2025
- Quantity
- 1 unit OUS
AlluraXper FD20 OR Table System Code: (1) 722023 (2) 722035
- Lot or code
- UDI: N/A
- FDA recall number
- Z-1083-2025
- Quantity
- 1 unit US; 1 unit OUS
AlluraXperFD20/10 System Code: (1) 722029
- Lot or code
- UDI: 00884838054219(21)
- FDA recall number
- Z-1084-2025
- Quantity
- 22 units US; 65 units OUS
Allura Xper FD20/20 System Code: (1) 722038
- Lot or code
- UDI: 00884838054226(21)
- FDA recall number
- Z-1085-2025
- Quantity
- 33 units US; 82 units OUS
AlluraXperFD20/15 System Code: (1) 722058
- Lot or code
- UDI: 00884838059115(21)
- FDA recall number
- Z-1086-2025
- Quantity
- 69 units US; 235 units OUS
Azurion 3 M12 System Code: (1) 722063 (2) 722221
- Lot or code
- UDI: (1) 722063 00884838085275(21) (2) 722221 00884838099203(21)
- FDA recall number
- Z-1087-2025
- Quantity
- 9 units US; 101 units OUS
Azurion 3 M15 System Code: (1) 722064 (2) 722222 (3) 722280
- Lot or code
- UDI: (1) 722064 00884838085282(21); (2) 722222 00884838099210(21); (3) 722280 00884838103276(21)
- FDA recall number
- Z-1088-2025
- Quantity
- 42 units US; 661 units OUS
Azurion 5 M12 System Code: (1) 722227 (2) 722231
- Lot or code
- UDI: (1) 722227 00884838099227(21); (2) 722231 00884838116740(21)
- FDA recall number
- Z-1089-2025
- Quantity
- 6 units US; 246 units OUS
Azurion 5 M20 System Code: (1) 722228 (2) 722232 (3) 722281
- Lot or code
- UDI: (1) 722228 00884838099234(21); (2) 722232 00884838116757(21); (3) 722281 00884838110564(21)
- FDA recall number
- Z-1090-2025
- Quantity
- 68 units US; 438 units OUS
Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235
- Lot or code
- UDI: (1) 722067 00884838085350(21); (2) 722225 00884838099265(21); (3) 722235 00884838116788(21)
- FDA recall number
- Z-1091-2025
- Quantity
- 90 units US; 458 units OUS
Azurion 7 B20 System Code: (1) 722068 (2) 722226 (3) 722236
- Lot or code
- UDI: (1) 722068 00884838085367(21); (2) 722226 00884838099272(21); (3) 722236 00884838116801(21)
- FDA recall number
- Z-1092-2025
- Quantity
- 251 units US; 938 units OUS
Azurion 7 M12 System Code: (1) 722078 (2) 722223 (3) 722233
- Lot or code
- UDI: (1) 722078 00884838085251(21); (2) 722223 00884838099241(21); (3) 722233 00884838116764(21)
- FDA recall number
- Z-1093-2025
- Quantity
- 324 units US; 1390 units OUS
Azurion 7 M20 System code: (1) 722224 (2) 722079 (3) 722234 (4) 722282 System Code: (1) 722078 (2) 722223 (3) 722233
- Lot or code
- UDI: (1) 722224 00884838099258(21); (2) 722079 00884838085268(21); (3) 722234 00884838116771(21); (4) 722282 00884838110571(21)
- FDA recall number
- Z-1094-2025
- Quantity
- 1346 units US; 2582 units
Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.