What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
4 more codes are in the record below.
. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function
- Lot or code
- Nobel Biocare N1 TiUltra TCC RP 4.0x 9mm Catalog Number: 300861 UDI-DI code: 07332747161731 Lot Numbers: 12253672 12253673 Nobel Biocare N1 TiUltra TCC RP 4.0x11mm Catalog Number: 300862 UDI-DI code: 07332747161748 Lot Numbers: 12253448 12254209 Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Catalog Number: 300863 UDI-DI code: 07332747161755 Lot Numbers: 12253620 12254213
- FDA recall number
- Z-1101-2025
- Quantity
- 1,558 units (90 units - U.S.)
Distribution (FDA): U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Nobel Biocare
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.