What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP3403
- Lot or code
- UDI-DI: 05055273204421 All lots within expiry
- FDA recall number
- Z-1017-2025
- Quantity
- 15 kits
Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP2757
- Lot or code
- UDI-DI: 05055273209198 All lots within expiry
- FDA recall number
- Z-1018-2025
- Quantity
- 33 kits
Distribution (FDA): US Nationwide distribution Including PR.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Randox Laboratories, Limited
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.