What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The Pin Guide is being recalled because the metal tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset being unable to detect the Pin Guide and potentially prolonging surgery time or changing of surgical method from guided to conventional.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty.
- Lot or code
- UDI/DI 03700434022926, Lot Numbers: 22A796, 22A797, 22A978, and 22A979
- FDA recall number
- Z-0846-2025
- Quantity
- 24 units
Distribution (FDA): US: MN, KY, MD, MA, TX, AK, TX, ID; France, Canada
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Tornier S.A.S.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.