What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential for signal drift in specific lots of BD PCR Cartridges.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
BD PCR Cartridges used with MAX System. Model Number: 437519
- Lot or code
- Model Number: 437519. UDI Numbers: (01)30382904375196(17)261205(10)4323345, (01)30382904375196(17)261129(10)4317725, (01)30382904375196(17)261122(10)4309022, (01)30382904375196(17)261115(10)4302135, (01)30382904375196(17)261111(10)4296086, (01)30382904375196(17)261104(10)4289593, (01)30382904375196(17)261029(10)4281532, (01)30382904375196(17)261021(10)4275828, (01)30382904375196(17)261014(10)4268747, (01)30382904375196(17)261007(10)4254714, (01)30382904375196(17)260930(10)4254710, (01)30382904375196(17)260923(10)4247089, (01)30382904375196(17)260913(10)4239079, (01)30382904375196(17)260909(10)4233297, (01)30382904375196(17)260902(10)4225364, (01)30382904375196(17)260826(10)4204545, (01)30382904375196(17)260805(10)4198065, (01)30382904375196(17)260729(10)4191581, (01)30382904375196(17)260722(10)4185634, (01)30382904375196(17)260716(10)4178328, (01)30382904375196(17)260704(10)4164798, (01)30382904375196(17)260701(10)4164796, (01)30382904375196(17)260701(10)4164792, (01)30382904375196(17)260701(10)4164794, (01)30382904375196(17)260701(10)4164789, (01)30382904375196(17)260624(10)4149773, (01)30382904375196(17)260617(10)4149768, (01)30382904375196(17)260610(10)4143963, (01)30382904375196(17)260605(10)4142324, (01)30382904375196(17)260604(10)4136790, (01)30382904375196(17)260528(10)4130663,(01)30382904375196(17)260527(10)4130589, (01)30382904375196(17)260520(10)4121669, (01)30382904375196(17)260517(10)4117376, (01)30382904375196(17)260513(10)4114952,…
- FDA recall number
- Z-1004-2025
- Quantity
- 516,435 units
The text above is the beginning of the record; 123 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Canada, South Korea, Taiwan, Australia, Singapore, Mexico, Hong Kong, India, Japan, Columbia, Brazil, Philippines, Argentina, New , Zealand, Uruguay, Chile, Trinidad and Tobago, Belgium, Vietnam, Macao.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Becton Dickinson & Co.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.