What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
There is a potential safety risk caused by physical disassociation of the back section from the bed frame of Skytron model series 3600, 6500, 6600 and 6700 operating room beds when used with the EnVisio Navigation System.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
EnVisio Patient Pad Transition Cushion, Model/Catalog Number: ENVISIO-SYS Product Description: The EnVisio Patient Pad Transition Cushion is a component of the EnVisio Navigation System which is a medical device that detects the presence of the SmartClip Soft Tissue Marker(s) and provides real-time three-dimensional navigation during surgery.
- Lot or code
- Lot Code: All EnVisio Patient Pad Transition Cushions REF: ENVISIO-CUSHION
- FDA recall number
- Z-0806-2025
- Quantity
- 283 (48 units)
Distribution (FDA): US Distribution: AZ, CA, CO, CT, FL, GA, IL, KY, MN, NC, ND, NJ, NY, OH, PA, SC, WI
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Elucent Medical Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.