What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
- Lot or code
- Model Number: LVP-0004, UDI Number: 00811505030320. Serial Numbers: 9233532106, 9233421830, 9233421828, 9233532158, 9233522076, 9233522041, 9233421833, 9233421829, 9233421808, 9233421654, 9233421626, 9233421621, 9233321340, 2215401517, 2215401545, 9233421789, 9233421809, 9233532167, 9233532100, 9233522070, 9233421825, 9233421824, 9233421794, 9233421783, 9233421781, 9233421771, 9233421768, 9233421756, 9233421736, 9233421713, 9233421702, 9233421617, 9233421611, 9233391513, 9233391491, 9233532098, 9233421822, 9233421797, 9233421678, 9233421666, 9233421601, 9233532114, 9233421792, 9233421766, 9233421687, 2215401372, 9233522040, 9233522037, 9233421764, 9233421752, 9233421665, 9232490602, 9233522062, 9233421817, 9233421832, 9233522094, 9233421805, 9233421740, 9233532161, 9233532111, 9233522061, 9233421836, 2212200949, 2212200956, 2212200959, 2212200960, 2212200964, 2212200965, 2212200983, 2212200999, 2212201000, 2212201001, 2212201006, 2212201008, 2212201009, 2212201011, 2212201017, 2212201020, 2212201021, 2212201035, 2212201037, 2212201038, 2212201043, 2215401572, 2219200223, 2219200257, 2219200268, 2219200273, 2219200304, 2219200317, 2219200381, 2223703322, 2223703362, 2223703372, 2223703378, 2223703388, 2223703421, 2223703437, 2223703461, 2223703484, 2223703490, 2223703494, 2223703508, 2223703523, 2223703534, 2223703541, 2223703547, 2318600285, 2318600382, 2318600387, 2321600019, 2321600020, 2321600022, 2321600023, 2321600026, 2321600027, 2321600039, 2321600046, 2321600048,…
- FDA recall number
- Z-0763-2025
- Quantity
- 1,520 units
The text above is the beginning of the record; 16,812 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): US distribution to states of: ID, MN, NV, TX, VA, WI.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Fresenius Kabi USA, LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.