What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Completed. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The extension sets subject to this recall have ENFit connections, although they should have non- ENFit connections. ENFit connectors will only fit with other ENFit-compatible devices, and will not fit with non-ENFit connectors.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
AMT Mini Enteral Extension Set, 12" Right Angle Feeding Set with Y-Port. Model/Catalog Number: E6-1222
- Lot or code
- Lot Code: Model No: E6-1222 UDI-DI: (01)00842071107086(17)270401(10)240424-628 Lot number: 240424-628 Exp. Date: 2027-04-01
- FDA recall number
- Z-0819-2025
- Quantity
- 100
Distribution (FDA): US Distribution to states of: CA, CT, FL, GA, IL, MA, MD, MO, NC, NJ, NY, OH, OK, PA, SC, TX, VA, WA
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Applied Medical Technology Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.