Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 95856: Stryker Corporation: 100V NEPTUNE 3 ROVER JAPANESE

FDA recall 95856 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

24 more codes are in the record below.

  • 100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V NEPTUNE 3 ROVER SPANISH (0703-002-000ES)

    Lot or code
    0703 002 000ENG GTIN 07613327382082, Lot Numbers; 2427418723 2427423863 2427418453 2427423873 2427418743 2427419773; 0703 002 000ES GTIN 07613327382068, Lot Numbers: 2427418453; 0703 003 000 GTIN 07613327381948, Lot Numbers: 2427418093 2427418053 2427418073 2427418233 2427418013 2427418723; 0711 001 000 GTIN 07613327576887, Lot Numbers: 2429720623 2429721033 2429720903 2429721203 2429720973 2429721223 2429721023 2303501403 2323610223 2403600253;
    FDA recall number
    Z-0807-2025
    Quantity
    14 (Expanded to 21 units on 10/9/25)

Distribution (FDA): US (Domestic) distribution to states of: WA, GA, and IL. OUS (International) distribution to Japan and Netherlands.

Source

Issued by
U.S. Food and Drug Administration
Firm
Stryker Corporation
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Stryker Corporation, Month: January 2025, Week 2025-W03. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95856 on January 15, 2025.

Other notices of Stryker Corporation