What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
17 more codes are in the record below.
DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens
- Lot or code
- UDI/DI 80567713190605F, Lot Numbers: 225084, 232094, 234114, 259144, 221214, 252244, 230094, 224124, 205144, 228174, 228224, 223274, 219104, 217134, 233154, 236174, 224244, 232294, 229224, 251234, 221294, 210204, 223244, 222314
- FDA recall number
- Z-0677-2025
- Quantity
- 21905 units
Distribution (FDA): Worldwide
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Diasorin Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.