What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
4 more codes are in the record below.
Sheath, 20 Fr., 130 mm (REF: WA2PS20S), gastroenterology-urology endoscopic access overtube.
- Lot or code
- REF: WA2PS20S; UDI-DI: 04251303810926; Lot Numbers: 154125
- FDA recall number
- Z-0713-2025
- Quantity
- 34 units
Sheath, 20 Fr., 160 mm (REF: WA2PS20L), gastroenterology-urology endoscopic access overtube.
- Lot or code
- REF: WA2PS20L; UDI-DI: 04251303810919; Lot Numbers: 154236
- FDA recall number
- Z-0714-2025
- Quantity
- 42 units
Sheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.
- Lot or code
- REF: WA2PS18S; UDI-DI: 04251303810865; Lot Numbers: 159825
- FDA recall number
- Z-0715-2025
- Quantity
- 53 units
Sheath, 18 Fr., 160 mm (REF: WA2PS18L), gastroenterology-urology endoscopic access overtube.
- Lot or code
- REF: WA2PS18L; UDI-DI: 04251303810872; Lot Numbers: 154214
- FDA recall number
- Z-0716-2025
- Quantity
- 102 units
Distribution (FDA): US (Domestic) distribution in state of PA and OUS (foreign) to country of Germany. .
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Trokamed GmbH
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.