What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
a. A small number of Fusion Pro 24 units were released with an incorrectly placed Interlock device. As a result, the system can be operated with the front-loading door in an open position. b. Fails to comply with Class 2 designation, indicated in Part 6 of Laser Product Report. An incorrectly placed Interlock device can allow Class 4 laser radiation fields during operation.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Fusion Pro 24, Model 17000
- Lot or code
- Model 17000 Product Report Accession Number: 2021046-000
- FDA recall number
- Z-0524-2025
- Quantity
- 751
Distribution (FDA): US Nationwide Distribution
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Epilog Laser Corp.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.