FDA recall 95729: Becton, Dickinson and Company, BD Bio Sciences: FACSLyric 2L4C Instrument US and 4 more

FDA recall 95729 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

140 more codes are in the record below.

  • FACSLyric 2L4C Instrument US, Catalog Number 662875, UDI 00382906628756 BD FACSLyric" flow cytometer is intended for in vitro diagnostic for the following: " Immunophenotyping using up to six fluorescence detection channels and two light scatter channels using a blue (488-nm) and a red (640-nm) laser. " It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.

    Lot or code
    Power supply part number 650781 with serial numbers beginning with V2309, V2318, and V2330 are impacted Catalog Number: 662875 UDI-DI code: 00382906628756 Instrument Serial Numbers/ Power Supply Serial Numbers: Z662875000119 V23300114 Z662875000117 V23300192 Z662875000112 V23300098 Z662875000123 V23300321 Z662875000105 V23090179 Z662875000104 V23090157 Z662875000107 V23090507 Z662875000102 V23090029 Z662875000103 V23090114 Z662875000106 V23090177 Z662875000108 V23180127 Z662875000110 V23300074 Z662875000111 V23090430 Z662875000109 V23300294 Z662875000116 V23300051
    FDA recall number
    Z-0745-2025
    Quantity
    15 power supplies
  • Device Description: FACSLyric 2L6C Instrument US, Catalog Number 662876, UDI 00382906628763 The K170974 clearance document describes the device as follows: The BD FACSLyric" flow cytometer (3-1, 4-2, 4-2-2, and 4-3-3 optical configurations) systems consist of a flow cytometer, sheath tank, waste tank, and a computer workstation. System options include an automated FACS Universal Loader and a barcode reader. The BD FACSLyric Flow Cytometer includes the 488 nm laser and 640 nm laser as part of four available manufactured instrument configurations. BD FACSLyric Flow Cytometer, 3-1 configuration, 4-color/2-laser BD FACSLyric Flow Cytometer, 4-2 configuration, 6-color/2-laser BD FACSLyric Flow Cytometer, 4-2-2 configuration, 8-color/3-laser BD FACSLyric Flow Cytometer, 4-3-3 configuration, 10-color/3-laser The lower level configurations are upgradeable to higher level configurations by adding filters, photomultiplier tubes (PMTs), and a laser. Only the 488 nm laser and 640 nm lasers are utilized for cleared in vitro diagnostic (IVD) applications and only fluorescence channels 1 (FL1) through FL6 are the subject of this 510(k) submission. Seven to tencolor immunophenotyping is for research use only (RUO). All optical configurations of the FACSLyric share the same dimensions: 22.8 inches in height by 24.93 inches in width by 22.8 inches in depth. The K201814 clearance document describes the device as follows: Refer to K170974 for detailed description and functioning of the BD…

    Lot or code
    Power supply part number 650781 with serial numbers beginning with V2309, V2318, and V2330 are impacted. Catalog Number: 662876 UDI-DI code: 00382906628763 Instrument Serial Numbers / Power Supply Serial Numbers: Z662876000065 V23300165 Z662876000044 V23090245 Z662876000045 V23090338 Z662876000048 V23180201 Z662876000047 V23180258 Z662876000051 V23180476 Z662876000050 V23180466 Z662876000049 V23180113 Z662876000052 V23180415 Z662876000053 V23180190 Z662876000056 V23300354 Z662876000055 V23300347 Z662876000057 V23300249 Z662876000059 V23300174 Z662876000061 V23300274 Z662876000062 V23300132 Z662876000060 V23300286 Z662876000063 V23300163 Z662876000066 V23300232 Z662876000054 V23300408
    FDA recall number
    Z-0746-2025
    Quantity
    20 power supplies

    The text above is the beginning of the record; 873 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • FACSLyric 3L8C Instrument US-IVD, Catalog Number 662877, UDI 00382906628770 The K170974 clearance document describes the device as follows: The BD FACSLyric" flow cytometer (3-1, 4-2, 4-2-2, and 4-3-3 optical configurations) systems consist of a flow cytometer, sheath tank, waste tank, and a computer workstation. System options include an automated FACS Universal Loader and a barcode reader. The BD FACSLyric Flow Cytometer includes the 488 nm laser and 640 nm laser as part of four available manufactured instrument configurations. BD FACSLyric Flow Cytometer, 3-1 configuration, 4-color/2-laser BD FACSLyric Flow Cytometer, 4-2 configuration, 6-color/2-laser BD FACSLyric Flow Cytometer, 4-2-2 configuration, 8-color/3-laser BD FACSLyric Flow Cytometer, 4-3-3 configuration, 10-color/3-laser The lower level configurations are upgradeable to higher level configurations by adding filters, photomultiplier tubes (PMTs), and a laser. Only the 488 nm laser and 640 nm lasers are utilized for cleared in vitro diagnostic (IVD) applications and only fluorescence channels 1 (FL1) through FL6 are the subject of this 510(k) submission. Seven to tencolor immunophenotyping is for research use only (RUO). All optical configurations of the FACSLyric share the same dimensions: 22.8 inches in height by 24.93 inches in width by 22.8 inches in depth. The K201814 clearance document describes the device as follows: Refer to K170974 for detailed description and functioning of the BD FACSLyric flow…

    Lot or code
    Power supply part number 650781 with serial numbers beginning with V2309, V2318, and V2330 are impacted Catalog Number: 662877 UDI-DI code: 00382906628770 Instrument Serial Numbers / Power Supply Serial Numbers: Z662877000064 V23180197 Z662877000053 V23090166 Z662877000052 V23090305 Z662877000054 V23090341 Z662877000051 V23090060 Z662877000056 V23090037 Z662877000058 V23090446 Z662877000055 V23090190 Z662877000061 V23180242 Z662877000057 V23090456 Z662877000060 V23180454 Z662877000065 V23180453 Z662877000062 V23300158 Z662877000063 V23300110 Z662877000059 V23180441 Z662877000068 V23300012
    FDA recall number
    Z-0747-2025
    Quantity
    16 power supplies

    The text above is the beginning of the record; 858 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • FACSLyric 3L10C Instrument US-IVD, Catalog Number 662878, UDI 00382906628787 The K170974 clearance document describes the device as follows: The BD FACSLyric" flow cytometer (3-1, 4-2, 4-2-2, and 4-3-3 optical configurations) systems consist of a flow cytometer, sheath tank, waste tank, and a computer workstation. System options include an automated FACS Universal Loader and a barcode reader. The BD FACSLyric Flow Cytometer includes the 488 nm laser and 640 nm laser as part of four available manufactured instrument configurations. BD FACSLyric Flow Cytometer, 3-1 configuration, 4-color/2-laser BD FACSLyric Flow Cytometer, 4-2 configuration, 6-color/2-laser BD FACSLyric Flow Cytometer, 4-2-2 configuration, 8-color/3-laser BD FACSLyric Flow Cytometer, 4-3-3 configuration, 10-color/3-laser The lower level configurations are upgradeable to higher level configurations by adding filters, photomultiplier tubes (PMTs), and a laser. Only the 488 nm laser and 640 nm lasers are utilized for cleared in vitro diagnostic (IVD) applications and only fluorescence channels 1 (FL1) through FL6 are the subject of this 510(k) submission. Seven to tencolor immunophenotyping is for research use only (RUO). All optical configurations of the FACSLyric share the same dimensions: 22.8 inches in height by 24.93 inches in width by 22.8 inches in depth. The K201814 clearance document describes the device as follows: Refer to K170974 for detailed description and functioning of the BD FACSLyric flow…

    Lot or code
    Power supply part number 650781 with serial numbers beginning with V2309, V2318, and V2330 are impacted Catalog Number: 662878 UDI-DI code: 00382906628787 Instrument Serial Numbers / Power Supply Serial Numbers: Z662878000426 V23180200 Z662878000412 V23180272 Z662878000410 V23180312 Z662878000046 UNK Z662878000319 UNK Z662878000367 V23090032 Z662878000373 V23090334 Z662878000370 V23090030 Z662878000371 V23090076 Z662878000377 V23090329 Z662878000369 V23090021 Z662878000375 V23090010 Z662878000372 V23090176 Z662878000378 V23090048 Z662878000380 V23090136 Z662878000374 V23090354 Z662878000383 V23090365 Z662878000379 V23090361 Z662878000401 V23180013 Z662878000382 V23090191 Z662878000381 V23090196 Z662878000421 V23180001 Z662878000387 V23090396 Z662878000420 V23180150 Z662878000425 V23180449 Z662878000399 V23180007 Z662878000400 V23090381 Z662878000422 V23090460 Z662878000386 V23090409 Z662878000388 V23090374 Z662878000390 V23090408 Z662878000368 V23090036 Z662878000431 V23300473 Z662878000432 V23300472 Z662878000389 V23090387 Z662878000423 V23180014 Z662878000454 V23300221 Z662878000413 V23180117 Z662878000376 V23090122 Z662878000455 V23300323 Z662878000453 V23300273 Z662878000411 V23180096 Z662878000385 V23090172 Z662878000384 V23090499 Z662878000394 V23090394 Z662878000395 V23180408 Z662878000434 V23180118 Z662878000435 V23300149 Z662878000111 UNK
    FDA recall number
    Z-0748-2025
    Quantity
    49 power supplies

    The text above is the beginning of the record; 858 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • FACSLyric 3L12C Instrument RUO, Catalog Number 663518, UDI 00382906635181

    Lot or code
    Power supply part number 650781 with serial numbers beginning with V2309, V2318, and V2330 are impacted Catalog Number: 663518 UDI-DI code: 00382906635181 Instrument Serial Numbers / Power Supply Serial Numbers: MP6635180000004 V23180079 MP6635180000008 V23180205 MP6635180000005 V23180222 MP6635180000006 V23180230 Z663518000594 V23180062 Z663518000529 V23300356 Z663518000526 V23300128 Z663518000470 V23180355 Z663518000471 V23180343 Z663518000456 V23090485 Z663518000433 V23090505 Z663518000402 V23090220 Z663518000165 UNK Z663518000196 UNK Z663518000214 UNK Z663518000338 V23090066 Z663518000342 V23090294 Z663518000339 V23090178 Z663518000349 V23090224 Z663518000350 V23090127 Z663518000357 V23090035 Z663518000343 V23090118 Z663518000346 V23090121 Z663518000352 V23090319 Z663518000360 V23090242 Z663518000355 V23090002 Z663518000359 V23090351 Z663518000340 V23090283 Z663518000337 V23090300 Z663518000335 V23090301 Z663518000361 V23090352 Z663518000362 V23090113 Z663518000347 V23090077 Z663518000336 V23090304 Z663518000341 V23090292 Z663518000365 V23090235 Z663518000358 V23090328 Z663518000353 V23090371 Z663518000382 V23090360 Z663518000369 V23090085 Z663518000380 V23090169 Z663518000351 V23090025 Z663518000348 V23090171 Z663518000356 V23090273 Z663518000378 V23090090 Z663518000345 V23090018 Z663518000367 V23090226 Z663518000385 V23090369 Z663518000376 V23090356 Z663518000379 V23090041 Z663518000354 V23090359 Z663518000384 V23090490 Z663518000386 V23090262 Z663518000388 V23090162…
    FDA recall number
    Z-0749-2025
    Quantity
    139 power supplies

    The text above is the beginning of the record; 3,336 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, Arkansas, California, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Maryland, Massachusetts, Michigan, Minnesota, Nebraska, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Tennessee, Texas, Virginia & Wisconsin. The countries of China, Costa Rica, Singapore, Thailand.

Source

Issued by
U.S. Food and Drug Administration
Firm
Becton, Dickinson and Company, BD Bio Sciences
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Becton, Dickinson and Company, BD Bio Sciences, Month: January 2025, Week 2025-W01. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional…”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95729 on January 1, 2025.