What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
155 more codes are in the record below.
Ziehm Solo. Digital Mobile C-arm
- Lot or code
- UDI: EZIEZIEHSOLO1. Serial Numbers: 50652, 51087, 51265, 52387, 52889, 53255, 54259, 54820, 54821, 54823, 54824, 54825, 54826, 55053, 55066, 55073, 55074, 55075, 55076, 55145, 55219, 55220, 55221, 55222, 55223, 55224, 55225, 55226.
- FDA recall number
- Z-0573-2025
- Quantity
- 28
Ziehm Solo FD. Interventional fluoroscopic x-ray system
- Lot or code
- UDI: EZIEZIEHMSOLOFD1 Serial Numbers: 54301, 54772, 54773, 54774, 54792, 54793, 54794, 54795, 54796, 54797, 54798, 54851, 54852, 54855, 55020, 55021, 55114, 55115, 55116, 55117, 55118, 55119, 55120, 55121, 55212.
- FDA recall number
- Z-0574-2025
- Quantity
- 25
Ziehm Vision system, x-ray, tomography, computed.
- Lot or code
- UDI: EZIEZIEHMVISIONFD1 Serial Numbers: 9246, 90155, 90416.
- FDA recall number
- Z-0575-2025
- Quantity
- 3
Ziehm Vision FD. Interventional fluoroscopic x-ray system
- Lot or code
- UDI: EZIEZIEHMVISIONFD1651 Serial Numbers: 93412, 93443, 93451, 93452, 93461, 93483, 93485, 93486, 93503, 93619, 93620, 93621. 93622 93623 93628 93629 93660 93661 93662 93626 93625 93627
- FDA recall number
- Z-0576-2025
- Quantity
- 22
Ziehm Vision R. Iterventional fluoroscopic x-ray system
- Lot or code
- UDI: EZIEZIEHMVISIONR1. Serial Numbers: 10295, 10490, 10656, 10716, 10997.
- FDA recall number
- Z-0577-2025
- Quantity
- 5
Ziehm Vision RFD. Interventional fluoroscopic x-ray system
- Lot or code
- UDI: EZIEZIEHMVISIONRFD1. Serial Numbers: 20048, 20240, 23805, 21350, 23249, 23250, 23251, 23252, 23253, 23324, 23375, 23376, 23377, 23378, 23379, 23380, 23381, 23382, 23412, 23452, 23460, 23461, 23462, 23463, 23506, 23536, 23721, 23737, 23738, 23739.
- FDA recall number
- Z-0578-2025
- Quantity
- 30
Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray system
- Lot or code
- UDI: EZIEZIEHMVISIONRFD3D1. Serial Numbers: 21118, 21212, 22040, 22710, 22764, 23247, 23248, 23255, 23257, 23258, 23259, 23260, 23327, 23328, 23329, 23456, 23457, 23458, 23459, 23653, 23654, 23655, 23656, 23657, 23666, 23667, 23668, 23669, 23670, 23671, 23672, 23673, 23674, 23723, 23740, 23744, 23745, 23746, 23747, 23748, 23749, 23750, 23751, 23752, 23753, 23775, 23801, 23802, 23828, 23849.
- FDA recall number
- Z-0579-2025
- Quantity
- 50
Distribution (FDA): U.S. (including Puerto Rico).
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Orthoscan, Inc.
- Title