What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190
- Lot or code
- Model Number: GIF-1TH190; UDI: 04953170343360, 04953170437014; Serial Number Range: Before 2446103 or After 2500453
- FDA recall number
- Z-0620-2025
- Quantity
- 1,770 units
Distribution (FDA): US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Olympus Corporation of the Americas (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.