What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP
- Lot or code
- REF 14-504-17, UDI/DI 35021239002579, Lot Numbers: 0003340, 0003620, 0003616, 0003580, 0003542, 0003509, 0003579, 0003568, 0003599, 0003615, 0003589, 0003562, 0003601, 0003618, 0003583, 0003613, 0003270, 0003629, 0003293, 0003584, 0003487, 0003617, 0003280, 0003348, 0003470, 0003523, 0003607, 0003576, 0003422, 0003614, 0003633, 0003578, 0003630, 0003207, 0003438, 0003559, 0003570, 0003478, 0003533, 0003588, 0003433, 0003556, 0003561, 0003446, 0003284, 0003545, 0003461, 0003590, 0003398, 0003530, 0003539, 0003550, 0003213, 0003598, 0003623, 0003504, 0003548, 0003552, 0003631, 0003594, 0003625, 0003600, 0003378, 0003526, 0003574, 0003411, 0003471, 0003587, 0003596, 0003297, 0003298, 0003412, 0003424, 0003425, 0003517, 0003151, 0003163, 0003235, 0003347, 0003350, 0003382, 0003390, 0003416, 0003423, 0003603, 0003638, 0003641, 0003175, 0003214, 0003238, 0003253, 0003257, 0003343, 0003351, 0003388, 0003397, 0003428, 0003486, 0003488, 0003496, 0003498, 0003528, 0003541, 0003608, 0003610, 0003612, 0003639, 0003640, 0003121, 0003122, 0003123, 0003125, 0003126, 0003127, 0003128, 0003129, 0003131, 0003132, 0003133, 0003134, 0003135, 0003136, 0003137, 0003138, 0003141, 0003142, 0003143, 0003145, 0003146, 0003147, 0003149, 0003150, 0003153, 0003154, 0003155, 0003156, 0003157, 0003158, 0003159, 0003160, 0003164, 0003166, 0003167, 0003168, 0003170, 0003171, 0003172, 0003173, 0003174, 0003176, 0003177, 0003178, 0003179, 0003180, 0003181, 0003182, 0003183, 0003188, 0003190, 0003191, 0003192,…
- FDA recall number
- Z-0527-2025
- Quantity
- 142327 units
The text above is the beginning of the record; 2,401 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP
- Lot or code
- REF 14-504-68, UDI/DI 35021239002586, Lot Numbers: 0003393, 0003394, 0003357, 0003358, 0003267, 0003368, 0003634, 0003254, 0003369, 0003165, 0003184, 0003196, 0003204, 0003206, 0003226, 0003256, 0003300, 0003307, 0003308, 0003314, 0003322, 0003359, 0003360, 0003361, 0003362, 0003363, 0003364, 0003377, 0003380, 0003635, 0003643; REF 14-504-68JP, UDI/DI 35021239002586, Lot Numbers: 0003369, 0003377, 0003393, 0003394, 0003634, 0003635, 0003643
- FDA recall number
- Z-0528-2025
- Quantity
- 10,223 units
smiths medical PORTEX Tracheal Tube Exchange Guide, 10Ch x 70cm, REF 14-504-76
- Lot or code
- UDI/DI 35021239002593, Lot Numbers: 0003373, 0003456, 0003621, 0003374, 0003376, 0003375, 0003459, 0003391, 0003480, 0003549, 0003301, 0003124, 0003130, 0003139, 0003140, 0003144, 0003148, 0003152, 0003161, 0003169, 0003185, 0003205, 0003210, 0003222, 0003232, 0003236, 0003255, 0003276, 0003277, 0003315, 0003329, 0003339, 0003367, 0003370, 0003371, 0003372, 0003381, 0003395, 0003399, 0003403, 0003414, 0003449, 0003457, 0003469, 0003476, 0003477, 0003489, 0003492, 0003500, 0003573, 0003484, 0003619, 0003557, 0003510, 0003522, 0003597, 0003511, 0003622, 0003553, 0003581, 0003605, 0003529,
- FDA recall number
- Z-0529-2025
- Quantity
- 12282 units
Distribution (FDA): Worldwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Smiths Medical ASD, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.