Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 95605: Smiths Medical ASD, Inc.: smiths medical portex Venn Reusable Tracheal Tube Introducer and 2 more

FDA recall 95605 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP

    Lot or code
    REF 14-504-17, UDI/DI 35021239002579, Lot Numbers: 0003340, 0003620, 0003616, 0003580, 0003542, 0003509, 0003579, 0003568, 0003599, 0003615, 0003589, 0003562, 0003601, 0003618, 0003583, 0003613, 0003270, 0003629, 0003293, 0003584, 0003487, 0003617, 0003280, 0003348, 0003470, 0003523, 0003607, 0003576, 0003422, 0003614, 0003633, 0003578, 0003630, 0003207, 0003438, 0003559, 0003570, 0003478, 0003533, 0003588, 0003433, 0003556, 0003561, 0003446, 0003284, 0003545, 0003461, 0003590, 0003398, 0003530, 0003539, 0003550, 0003213, 0003598, 0003623, 0003504, 0003548, 0003552, 0003631, 0003594, 0003625, 0003600, 0003378, 0003526, 0003574, 0003411, 0003471, 0003587, 0003596, 0003297, 0003298, 0003412, 0003424, 0003425, 0003517, 0003151, 0003163, 0003235, 0003347, 0003350, 0003382, 0003390, 0003416, 0003423, 0003603, 0003638, 0003641, 0003175, 0003214, 0003238, 0003253, 0003257, 0003343, 0003351, 0003388, 0003397, 0003428, 0003486, 0003488, 0003496, 0003498, 0003528, 0003541, 0003608, 0003610, 0003612, 0003639, 0003640, 0003121, 0003122, 0003123, 0003125, 0003126, 0003127, 0003128, 0003129, 0003131, 0003132, 0003133, 0003134, 0003135, 0003136, 0003137, 0003138, 0003141, 0003142, 0003143, 0003145, 0003146, 0003147, 0003149, 0003150, 0003153, 0003154, 0003155, 0003156, 0003157, 0003158, 0003159, 0003160, 0003164, 0003166, 0003167, 0003168, 0003170, 0003171, 0003172, 0003173, 0003174, 0003176, 0003177, 0003178, 0003179, 0003180, 0003181, 0003182, 0003183, 0003188, 0003190, 0003191, 0003192,…
    FDA recall number
    Z-0527-2025
    Quantity
    142327 units

    The text above is the beginning of the record; 2,401 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP

    Lot or code
    REF 14-504-68, UDI/DI 35021239002586, Lot Numbers: 0003393, 0003394, 0003357, 0003358, 0003267, 0003368, 0003634, 0003254, 0003369, 0003165, 0003184, 0003196, 0003204, 0003206, 0003226, 0003256, 0003300, 0003307, 0003308, 0003314, 0003322, 0003359, 0003360, 0003361, 0003362, 0003363, 0003364, 0003377, 0003380, 0003635, 0003643; REF 14-504-68JP, UDI/DI 35021239002586, Lot Numbers: 0003369, 0003377, 0003393, 0003394, 0003634, 0003635, 0003643
    FDA recall number
    Z-0528-2025
    Quantity
    10,223 units
  • smiths medical PORTEX Tracheal Tube Exchange Guide, 10Ch x 70cm, REF 14-504-76

    Lot or code
    UDI/DI 35021239002593, Lot Numbers: 0003373, 0003456, 0003621, 0003374, 0003376, 0003375, 0003459, 0003391, 0003480, 0003549, 0003301, 0003124, 0003130, 0003139, 0003140, 0003144, 0003148, 0003152, 0003161, 0003169, 0003185, 0003205, 0003210, 0003222, 0003232, 0003236, 0003255, 0003276, 0003277, 0003315, 0003329, 0003339, 0003367, 0003370, 0003371, 0003372, 0003381, 0003395, 0003399, 0003403, 0003414, 0003449, 0003457, 0003469, 0003476, 0003477, 0003489, 0003492, 0003500, 0003573, 0003484, 0003619, 0003557, 0003510, 0003522, 0003597, 0003511, 0003622, 0003553, 0003581, 0003605, 0003529,
    FDA recall number
    Z-0529-2025
    Quantity
    12282 units

Distribution (FDA): Worldwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Smiths Medical ASD, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Smiths Medical ASD, Inc., Month: December 2024, Week 2024-W49. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95605 on December 4, 2024.

Other notices of Smiths Medical ASD, Inc.