FDA recall 95602: St. Jude Medical: CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics… and 1 more

FDA recall 95602 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

263 more codes are in the record below.

  • CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000

    Lot or code
    MN6000 (Which interfaces with CM1100 for PES) UDI-DI Code: 05414734510035 Merlin ID / PES Serial Number: 027556711 CM1100 P180912371 444058655 CM1100 P240100658 494955560 CM1100 P230300384 497887091 CM1100 P180208671 513914390 CM1100 P230800825 742148364 CM1100 P220800695 751954397 CM1100 P211200289 000697912 CM1100 P181013064 001545862 CM1100 P220400808,P220100189 003473584 CM1100 P211100353 005842000 CM1100 P230100314 006333918 CM1100 P220600729 008418758 CM1100 P240300492 010473700 CM1100 P210700575 010485845 CM1100 P220100057 011307774 CM1100 P190800198 013397427 CM1100 P220500363 018103333 CM1100 P210201417 018852830 CM1100 P190900022 019716455 CM1100 P221100752 020956348 CM1100 P230500703 021148283 CM1100 P160900736 021410246 CM1100 P230800171 023874951 CM1100 P230400238 024386088 CM1100 P200100300 026473926 CM1100 P240500531 031880776 CM1100 P240300135 032455370 CM1100 P240300138 034701771 CM1100 P230300580 037825221 CM1100 P240100040 039458344 CM1100 P190700191 042069666 CM1100 P221200307 043804384 CM1100 P220401019 044375798 CM1100 P200100451 046139531 CM1100 P210400699 046659165 CM1100 P220700471 049107915 CM1100 P220600696 049584410 CM1100 P230800848 053752093 CM1100 P230900101 053919908 CM1100 P200600006 054799366 CM1100 P211100584 055804058 CM1100 P180912650 056345895 CM1100 P210200735 057332710 CM1100 P240100823 058913997 CM1100 P240100775 060968161 CM1100 P240200495 061420626 CM1100 P230500071 062897822 CM1100 P240300373 066957044 CM1100 P210700838 067353847…
    FDA recall number
    Z-0988-2025
    Quantity
    605 units

    The text above is the beginning of the record; 15,558 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Hosptial Electronics System (PES) REF CM3000 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000

    Lot or code
    MN6000 (Which interfaces with CM3000 for HES) UDI-DI Code: 05414734510035 HES Serial Number/Clinic Merlin ID: M140400470 50011924 M140901162 50012084 M150102238 50013199 M150302655 50011863 M150905419 100351 M161107929 50012353 M161207976 50012166 M170108094 50011331 M170208159 50012146 M170208161 100268 M170408244 50015053 M170508275 50000752 M171108426 50051809 M171108437 50011840 M171108442 50011887 M180208509 50011887 M180308526 100710 M180608651 50011332 M180608652 50025010 M180608653 50015252 M180608669 50015252 M180608712 50060412 M180608727 50017544 M181008882 50038595 M190208990 50040301 M190208994 50040237 M190300010 50011866 M190400008 500150 M190700005 100605 M200400018 100359 M200600002 100351 M200600007 100080 M200800004 50050337 M200900014 50016848 M201100007 50050201 M210100003 9008126 M210300009 50014260 M210300011 50054572 M210300022 50013502 M210300023 50011879 M210900025 12000635 M211000003 50017405 M220600048 50025317
    FDA recall number
    Z-0989-2025
    Quantity
    431 units

Distribution (FDA): Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NE, NH, NJ, NV, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and the countries of Australia, and Canada.

Source

Issued by
U.S. Food and Drug Administration
Firm
St. Jude Medical
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: St. Jude Medical, Month: January 2025, Week 2025-W05. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95602 on January 29, 2025.

Other notices of St. Jude Medical