What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Within the Irish Market, a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life. There is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient. Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
4 more codes are in the record below.
Defibtech DDP-100 Adult Defibrillation Pads For use with AED Models: Lifeline and Lifeline Auto (DDU-1:XX Series)- Indicated for use on victims of sudden cardiac arrest (SCA)
- Lot or code
- UDI-DI: 00815098020423, DDP-100 adult defibrillation pads, single configuration; 30815098020424, DDP-100 adult defibrillation pads, 10 pairs; 50815098020428, DDP-100 adult defibrillation pads, 40 pairs;
- FDA recall number
- Z-0502-2025
- Quantity
- 22,294 total
Defibtech DDP-200P: Pediatric Defibrillation Pads 0 For use with AED Models: Lifeline and Lifeline Auto (DDU-l XX Series)-Indicated for use on victims of sudden cardiac arrest (SCA.)
- Lot or code
- UDI-DI: 00815098020430, DDP-200P pediatric defibrillation pads, single pair; 10815098020437, DDP-200P pediatric defibrillation pads, single pair configuration; 40815098020438, DDP-200P pediatric defibrillation pads, 20 pairs
- FDA recall number
- Z-0503-2025
- Quantity
- 22,294 total
Defibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, lifeline ECG, lifeline View (DDU-2XXX Series). -Indicated for use on victims of sudden cardiac arrest (SCA.)
- Lot or code
- UDI-DI: 00815098020447, DDP-2001 adult defibrillation pads, single pack; 30815098020448, DDP-2001 adult defibrillation pads, 10 pairs; 50815098020442, DDP-2001 adult defibrillation pads, 40 pairs
- FDA recall number
- Z-0504-2025
- Quantity
- 22,294 total
Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX Series -Indicated for use on victims of sudden cardiac arrest (SCA.)
- Lot or code
- UDI-DI: 00815098020454, DDP-2002, pediatric defibrillation pads, single pair; 10815098020451, DDP-2002, pediatric defibrillation pads, single pair configuration; 40815098020452, DDP-2002, pediatric defibrillation pads, 20 pairs
- FDA recall number
- Z-0505-2025
- Quantity
- 22,294 total
Distribution (FDA): Ireland
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Defibtech, LLC
- Title