Why it was recalled
Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
20 more codes are in the record below.
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4720
- Lot or code
- Catalog Number: 4720; UDI-DI: M84947200; Lot Numbers: H07-23-103, H07-23-104, H08-23-020, H09-23-111.
- FDA recall number
- Z-0468-2025
- Quantity
- 70 units
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4702
- Lot or code
- Catalog Number: 4702; UDI-DI: M84947020; Lot Numbers: H08-23-263, H01-24-114.
- FDA recall number
- Z-0469-2025
- Quantity
- 45 units
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4706
- Lot or code
- Catalog Number: 4706; UDI-DI: M84947060; Lot Number: SD09-23-089.
- FDA recall number
- Z-0470-2025
- Quantity
- 6 units
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4716
- Lot or code
- Catalog Number: 4716; UDI-DI: M84947160; Lot Number: H01-24-076.
- FDA recall number
- Z-0471-2025
- Quantity
- 5 units
Posi-Stop Injection Needle, 23 Gauge, Catalog Number: 4721
- Lot or code
- Catalog Number: 4721; UDI-DI: M84947210; Lot Number: H11-23-057.
- FDA recall number
- Z-0472-2025
- Quantity
- 15 units
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704
- Lot or code
- Catalog Number: 4704; UDI-DI: M84947040; Lot Numbers: H10-23-106, H11-23-059.
- FDA recall number
- Z-0473-2025
- Quantity
- 6 units
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4718
- Lot or code
- Catalog Number: 4718; UDI-DI: M84947180; Lot Numbers: H08-23-264, H09-23-112.
- FDA recall number
- Z-0474-2025
- Quantity
- 12 units
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4722
- Lot or code
- Catalog Number: 4722; UDI-DI: M84947220; Lot Numbers: H08-23-265, H09-23-113, H10-23-223, H11-23-058.
- FDA recall number
- Z-0475-2025
- Quantity
- 198 units
Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4723
- Lot or code
- Catalog Number: 4723; UDI-DI: M84947230; Lot Numbers: H07-23-107, H08-23-034.
- FDA recall number
- Z-0476-2025
- Quantity
- 10 units
Distribution (FDA): Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Hobbs Medical, Inc.
- Title