FDA recall 95502: Stryker Corporation: Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil and 5 more

FDA recall 95502 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

389 more codes are in the record below.

  • Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-000

    Lot or code
    UDI-DI: 07613327117202 Lot Numbers: 53863666 54409859 56319954 56945653 57699375 59053479 65178945 53863667 54409860 56319955 56945654 57699376 59053480 65178946 53863668 54409862 56319956 57699353 57699377 59053481 65178947 53863669 54409919 56319957 57699354 57699378 59053482 65178949 53863670 54409920 56319958 57699355 57699379 59053483 65347997 53863671 54409921 56319959 57699359 57699380 61288937 65784636 54403959 54409922 56319960 57699360 57699381 61288943 65784637 54403960 54409923 56319961 57699361 57699382 62825781 65784638 54403961 54409924 56892676 57699362 57805609 64117282 65784639 54403962 54409925 56892677 57699363 57805610 64477143 65876797 54403963 54409926 56892689 57699364 57805611 64477144 65961918 54403964 54409927 56932527 57699365 58320366 64477145 66133414 54403965 54533880 56932528 57699366 58514707 64855504 66275138 54409851 54533881 56932533 57699367 58766534 64855505 V53051139 54409852 54533882 56932534 57699368 58766535 64855515 V53051140 54409853 54533883 56945645 57699369 58766536 64855516 V53267506 54409854 54533884 56945646 57699370 58766539 64973261 V53526100 54409855 54654973 56945647 57699371 58766540 64973262 V53708277 54409856 54665296 56945648 57699372 58791419 64973263 V53708278 54409857 56319952 56945651 57699373 59053477 64973266 V53934891 54409858 56319953 56945652 57699374 59053478 64973267 V53934892 V53934893 Additional Lot Numbers: 64262566 66387346 66599251 66758654 67377504 67652611 67794514 64973268 66387347 66705549 66758655…
    FDA recall number
    Z-0365-2025
    Quantity
    1,120,860 eaches

    The text above is the beginning of the record; 216 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Rocker-Switch, Bulk Non-Sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-002

    Lot or code
    GTIN: 07613327379471 Lot Numbers: 53700305 53864207 53864241 53864247 60308905 62012084 64112630 65342955 53700306 53864212 53864243 53864248 60308906 62291715 64112631 65869470 53700307 53864213 53864244 53864249 60308907 62291717 64142443 65869471 53700308 53864214 53864245 53864257 61283868 62821201 64968515 65954806 53864205 53864219 53864246 60308904 61283869 64112629 64968516 66005707 53864206 Additional Lot Numbers: 64917537 65342956 66487443 67787272 67840867 68672644 64917538 66487442 67787271 67840866 67840868 68672645
    FDA recall number
    Z-0366-2025
    Quantity
    295,760 eaches
  • Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Non-coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-001

    Lot or code
    GTIN: 07613327117219 Lot Number: 64117291
    FDA recall number
    Z-0367-2025
    Quantity
    2,620 eaches
  • Stryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-000

    Lot or code
    GTIN: 07613327117202 Lot Numbers: 54409778 59692168 62016967 63292654 64324200 64855491 65961907 67325428 57699286 60313543 62016978 63292655 64324201 64855498 65961908 67377503 57738329 60313544 62296666 63632847 64324207 64855499 65961909 67481708 57738330 60313545 62296667 63632848 64324208 64855500 66012917 67481709 57738331 60313546 62296668 63632851 64354844 64855501 66012918 67481710 57738332 60313547 62382103 63632852 64354845 64973248 66133412 67481711 57738333 60313549 62382104 63842616 64354846 64973249 66387344 67481712 57738334 60834119 62382105 63842618 64354847 64973253 66494921 67652608 58014507 60834120 62678399 63842619 64354850 64973254 66494922 67652609 58014508 60834121 62678400 63842620 64354851 64973255 66494923 67722909 58014509 60834122 62678401 63871520 64477131 64973256 66494924 67794511 58014519 60834123 62678402 63871521 64477132 64973257 66494925 V56951192 58514679 60834124 62678403 63871522 64477135 64973258 66599248 V56951193 58514680 60834125 62678404 63871523 64477136 64973259 66599249 V56951194 59053422 61288888 63090872 63871524 64573557 64973260 66705547 V56951195 59053423 61288889 63090873 64117262 64573561 65178940 66705548 V56951196 59417841 61288890 63090874 64117263 64573562 65178942 66758641 V56951197 59417842 61288891 63090875 64117264 64573563 65178943 66758642 V58582358 59417843 61288892 63090876 64117265 64573564 65178944 66758643 V58582359 59417844 61288893 63090877 64117266 64604782 65784634 66758644 V58582360 59417845 61288895…
    FDA recall number
    Z-0368-2025
    Quantity
    1,996,010 eaches

    The text above is the beginning of the record; 390 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-001

    Lot or code
    GTIN: 07613327117196 Lot Numbers: 53867671 54409847 59053476 64855502 66494932
    FDA recall number
    Z-0369-2025
    Quantity
    21,490 eaches
  • Stryker Neptune SafeAir Smoke Evacuation Pencil Coated Push-Button, Bulk Non-sterile - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect. Catalog Number: 0703-046-002

    Lot or code
    GTIN: 07613327379488 Lot Numbers: 53864143 57458172 58761873 63837701 64258041 65777003 66126554 66751061 68311144 54406823 57458173 58761874 63837702 64258042 65777004 66487383 67200500 68311145 54406824 57458174 60308898 63866394 64258044 65954742 66487384 67200501 68672590 54406825 57458175 60308899 63866395 64850548 65954743 66487385 67200502 68728239 54406826 57694644 62291709 64052771 64850549 65954744 66487386 67715714 68728240 54406827 57733684 62377382 64112622 64968511 65954745 66751059 67787217 54406828 58009712 62537206 64112623 64968512 66126553 66751060 68200024
    FDA recall number
    Z-0370-2025
    Quantity
    403,730 eaches

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.

Source

Issued by
U.S. Food and Drug Administration
Firm
Stryker Corporation
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Stryker Corporation, Month: November 2024, Week 2024-W47. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 95502 on November 20, 2024.

Other notices of Stryker Corporation